Geiger-Müller Counter   513610 

 

AC513610  

 p

. 7/8 

 

 

Cancel  

You  can  exit  navigation  mode  by  "travelling  too 
far": Press  

  when

 Time

 is selected – or press  

   

when 

Ready Beep

 is selected. Pressing  

  twice 

gives  the  same  result  -  you  choose  to  set  a 
parameter, and accept the old, unchanged value. 

If  you  have  changed  a  parameter,  there  is  no 
"escape  button”.  However,  apart  from  the  GM-
voltage, everything can be reset by turning off and 
on again. The GM voltage should be set at 500 V 
when using Frederick's GM tubes.  

Battery Saver Function  

The unit turns off automatically after an hour - but 
only if it is inactive.  

 

 

 

Communication with a computer 

Communication  requires  a  cable  with  built-in 
adapter (512565). The adapter has an associated 
driver  that  must  be  installed  on  your  computer. 
Find information about downloading this driver on 
the packaging for the adapter or on our website. 
The adapter is recognized as a 

virtual serial port

Communication  parameters  can  be  set  in 
Windows’ Control Panel. 

Communication protocol 

Commands  to  the  unit  consist  of  one  letter, 
eventually 

followed 

by 

number. 

The possible responses from the device depend on 
the command. 

C: Read Company Information 

Returns  the  string  "(c)  Copyright  2010  A/S  S. 
Frederiksen" 

D: Read register values 

Returns  the  current  values  of  secondary  and 
primary data registers 

E: Send results or not 

When  used  alone,  it  returns  status.  When  used 
with  parameter  0-1,  the  status  will  change 
according to the following: 

'e0' 

 counts not automatically sent 

'e1' 

 counts sent automatically when each 

count period has finished 

 

F: Counting time 

When  used  alone,  it  returns  the  actual  counting 
time in seconds. When used with parameter 0-5, 
counting  period  is  changed  according  to  the 
parameter: 

'f0' 

  Infinitely 

'f1' 

  1s 

'f2' 

  10s 

'f3' 

  60s 

'f4' 

  100s 

'f5' 

  300s 

Other  counting  periods  are  invalid.  Display  icons 
are updated accordingly. 

J: GM voltage 

When  used  alone,  it  returns  the  current  GM 
voltage in volts. When used with parameter 400-
700, the GM voltage is modified according to the 
parameter. Example: 

'j520' 

 GM voltage set to 520V 

O: Single / Repeat Mode 

When  used  alone,  it  returns  status.  When  used 
with parameter 0-1, mode is changed accordingly: 

'o0' 

 Single 

'o1' 

 Repeat 

 

Summary of Contents for 513610

Page 1: ...den praktisk i repeat mode Fleksibel med hensyn til det anvendte r r Kan styries fra PC Mulighed for digitale pulser ud Specialkabler ekstraudstyr Quick guide Tilslut T lleren accepterer enten r r med...

Page 2: ...il en igangv rende m ling forts tte efter tryk p indtil den nye m leperiode udl ber Dette vil kikse hvis der v lges en tid som allerede er udl bet Hvis m leresultatet p denne m de g res ugyldigt nulst...

Page 3: ...tet best r af et bogstav evt efterfulgt af et tal De mulige responser fra apparatet afh nger af kommandoen C L s Firmaoplysninger Returner strengen c copyright 2010 A S S Frederiksen D L s registerv r...

Page 4: ...andoen alene returneres den aktuelle h jttaler status Bruges kommandoen med et tal vil h jttaleren tilsluttes afbrydes i h t tallet U0 GM sound fra Ready Beep fra U1 GM sound til Ready Beep fra U2 GM...

Page 5: ...ially handy in repeat mode Very flexible regarding the GM tubes used May be controlled from a PC Digital pulse output Special cables required option Quick Guide Connect The counter accepts tubes with...

Page 6: ...new period expires This will fail when you choose a time that has already expired If the measurement in this way is invalid the display will be reset When a measurement with fixed time is under way t...

Page 7: ...tion protocol Commands to the unit consist of one letter eventually followed by a number The possible responses from the device depend on the command C Read Company Information Returns the string c Co...

Page 8: ...ead If the register is empty when this command is received a value of 1 is returned To ensure that all values are transferred to the PC this register should be read on intervals not longer than the co...