PREFACE

Electrochemical water disinfection a safe and effective way to help sanitize the sink and help keep the drain and trap free of 

CPOs & other pathogens. During use and at periodic intervals, an electrolytic process occurs which reconstitutes the water 

molecules into mixed oxidants which are hostile to bacteria. No chemicals or consumables are added or used during this 

process. During use this electrolyzed water will flood the drain and aid in cleaning and disinfection. Please note this process 

is intended to augment (not replace) current cleaning or infection control protocols at the facility. 

Other features include a standard 20 second lathering timer and a post-wash cycle of 3 seconds of water runtime which 

helps wash all residue down into the sink drain. A daily purge cycle is also standard to help prevent stagnant water in the 

lines, maintain outlet temperatures and ensure continuous waste treatment. Medi-flo is designed for hand washing only 

and is not to be used as a drinking source, to wash instruments or bedpans etc. Recommended for areas with normal water 

conductivity above 75 TDS (total dissolved solids). If unaware of your local water conductivity, please contact customer 

service and this can easily be determined for you before you install the product. 

Medi-flo is recommended to be installed with a thermostatic mixing valve and set to a temperature of 27-38C. (Optional 

accessory MEDI-MIX is available from Franke.) Franke Medi-flo sink is designed to install on provided wall hanger brackets. If 

mounting on a wall that cannot support the load, we recommend the use of our FRANKE in-wall carrier IWC2104.

15 1/4 [387 mm]

31 9/16 [801 mm]

*

32 13/16 [833 mm]

*

2 [51 mm]

4 [102 mm]

Hot and Cold Supply Lines

Wall Hanger Brackets (2)

Secure to front of finished wall.

4 15/16 [125 mm]

Finished

Wall

Finished Floor

Integral

Carrier 

Rough-in 

Box

Notes:

1. * Yields a finished rim height of 33" (832mm)

2.  Recommended carrier FRANKE  IWC2104

1 3/4" (45mm) out 

from finished wall

2 7/8 [73 mm]

Optional Thermostatic Mixing Valve 

(MEDI-MIX)

Electronic Solenoid

Shroud Mount





Œ

Œ

EN

Summary of Contents for HWSS2321W-00

Page 1: ...INFECTION CONTROL WASH BASIN ...

Page 2: ...ste treatment Medi flo is designed for hand washing only and is not to be used as a drinking source to wash instruments or bedpans etc Recommended for areas with normal water conductivity above 75 TDS total dissolved solids If unaware of your local water conductivity please contact customer service and this can easily be determined for you before you install the product Medi flo is recommended to ...

Page 3: ... Location 17 5 8 448 mm 1 1 2 Waste Location FINISHED FLOOR Ø1 Ø25 mm 2 51 mm 3 1 2 89 mm 7 178 mm 31 9 16 801 mm 3 1 4 83 mm 3 7 8 98 mm WALL OPENINGS LAYOUT 21 1 4 539 mm 8 11 16 220 mm 10 3 8 263 mm Opening for rough in box 32 13 16 833 mm 2 2 2 2 EN No tools Hand tighten ...

Page 4: ...pace is limited to install the shroud We recommend a 1 piece tubular trap without a cleanout plug on the bottom the trap 5 5 ELECTRICAL CONNECTIONS Check local codes for compliance prior to making any electrical connections The water tight fittings have an alignment feature Gently push the fittings together rotating until they engage Little to no force should be required to do so Once fitted toget...

Page 5: ... as follows IR sensor will see a users hand This will trigger the water flow LED and the electrolytic generator will turn on When the user removes their hands to lather the water flow will turn off however the electrolytic generator will stay on for 20 seconds unless programmed otherwise The electrolytic generator is self programming based on local water conductivity If conductivity is in normal r...

Page 6: ...trolytic generator is not seen running 1 Check cable connection 2 Ensure water is not distilled or R O filtered TDS 50 Water output is too high 1 Adjust shut off valve installed in 3 8 line 2 Faulty flow control regulator found in brass output port of the solenoid valve Water output is too low 1 Faulty flow control regulator found in brass output port of the solenoid valve 2 Obstruction on dirt st...

Page 7: ... or by sending in a sample of 250ml minimum to our customer service Tests such as ORP and Indigo may not give valid readings due to hydrogen interference temperature water composition and PH Subtract the reading obtained from straight tap water to remove any reading from chlorine The default operation of the electrolytic generator is to run a charge cycle after every use Target canister charge is ...

Page 8: ... 00 Medizmi102218 V15 EN EN WARRANTY If inspection the product confirms that it is defective in materials or workmanship Franke Kindred Canada Ltd will at their choosing repair or exchange the product during the following periods Sink waste fitting and shroud 2 years Water controls and electronics 1 year NOTE In the event of a limited warranty claim proof of purchase date will be required ...