Page
4
MT04-001 AquariTherm Tap (Manual)
www.dartvalley.co.uk
Page
5
MT04-001 AquariTherm Tap (Manual)
www.dartvalley.co.uk
Re
vi
si
on N
o. M
AQ
TI
N
16
07
13
01
Re
vi
si
on N
o. M
AQ
TI
N
16
07
13
01
Step 9:
Continued...
1. Using a 4mm slotted screwdriver, insert through the hole
and locate the setting screw (Fig 11).
2. Turn the setting screw clockwise to reduce the
temperature and counter clockwise to increase the
temperature.
3. Only adjust by a small amount, a 1/4 turn in either
direction should be sufficient to reach the desired
temperature.
4. When desired temperature has been set, replace the
spindle plug.
Refit the penny washer, retainer screw and the screw cover.
Fig 11
Step 10:
Maintenance and Servicing
Step 6 :
Mounting the Spigot and Plumbing
Step 5 :
Typical Schematic
Page
6
Page
3
Fig 12
Fig 14
Fig 15
Fig 13
Mount the spigot to the wall surface and secure with the nut
and washer provided (Fig 3).
Connect the HOT and COLD water supply to the spigot using the
copper tails provided with the kit (Fig4) taking care to ensure the
HOT and COLD supply are connected correctly as indicated by
the coloured indicators. It is important that all pipework should be
thoroughly flushed through before connecting the tap.
Check ball valves on wall spigot are in the closed position as
Fig. 2.
Fig 2
Fig 3
Fig 4
HOT
COLD
INCREASE
MAX TEMPERATURE
The thermostatic cartridge should give many years of trouble –free
service, but in the event of failure or where routine maintenance is
required servicing is straightforward.
10.1 Filter and Check Valve
To service the internal filter and check valve the main tap body
needs to be removed from the spigot.
Using a 6mm slotted screw driver isolate the hot water supply by
turning the integrated isolation valve 90 clockwise (Fig 7.1).
Follow the same procedure for the cold supply.
Remove the retainer screw (Fig. 6 and 6.1).
The main tap body can now be removed to enable servicing of the
filter and check valve.
The filter cartridges (Fig 13) can be removed from the body of the
tap (Fig 12).
The filters can be removed and replaced. Service kits are
available, contact [email protected] or call (01803) 529021.
10.2 Thermostatic Cartridge
To remove the thermostatic cartridge, isolate water and power
supply, then remove screw cover and unscrew the retaining screw
and remove penny washer, top cap and limit ring (Fig 9). Unscrew
the cartridge retaining nut and pull the cartridge out. Clean or
replace cartridge as necessary. Be sure to re-fit the cartridge
with its locating lug pointing forwards (Fig 14). Refit the cartridge
retaining nut and tighten to 35Nm (Fig 15). Without the limit ring
fitted, place the Top Cap in position. Reactivate the cold and hot
water supplies.
Gently turn the Top Cap clockwise until the water just stops
flowing, then turn counterclockwise through approximately 90
degrees. Remove the Top Cap and position the limit ring on the
cartridge retaining nut, as shown (Fig 16).
Refit the Top Cap and secure the with retainer screw and penny
washer. Refit screw cover so that the logo is square.
Fig 16
LIMIT STOPS
½ TURN (35NM)
REDUCE
MAX TEMPERATURE
3 AMP FUSED SPUR
POWER SUPPLY
COLD WATER
SUPPLY
HOT WATER
SUPPLY
ISOLATOR
FILTER
SOLENOID
VALVE
Y-SPLITTER
LEAD
DUAL VALVE
ACTUATOR
AQUARITHERM
TAP
PROGRAMMABLE
SENSOR
Please reference AC02-017 instructions for sensor operation.
Available on www.dartvalley.co.uk