6. Wipe X Galvo Mirror

Prepare a new swab. It is 

important 

to use a fresh swab so as not to spread 

any particles or residues that had been caught by the previous swab.

Carefully position the galvo block as shown. Use the swab to gently rotate the 

mirror so that the surface is more easily accessible. Gently wipe the mirror 

from the base of the mirror to the end of the mirror. Rotate the swab and 

wipe again.

.

Once the mirrors are clean, the galvo block can be reinstalled into the Form 

1+. Use a bright light or flashlight to illuminate the mirrors. 

7. Inspect the Mirrors

6/7

DO NOT COPY OR REDISTRIBUTE THIS INFORMATION

Summary of Contents for GALVO

Page 1: ...ace mirrors in the Form 1 can cause scratches and irreversible damage Exercise extreme caution when the mirror is exposed Contact support formlabs com with any questions Formlabs Customer Support can...

Page 2: ...screws under the cover Some machines may also have two silver countersunk screws along the rear edge on the bottom of the Form 1 Carefully remove the back panel after all screws have been removed Prop...

Page 3: ...m the Form 1 frame WARNING DO NOT REMOVE OR TAMPER WITH ANY OTHER SCREWS IN THE GALVO BLOCK ADJUSTING ANY OTHER SCREWS WILL MOVE THE PRINTER OUT OF CALIBRATION Cut a strip about 1cm wide out of PEC PA...

Page 4: ...PAD over to the side Begin wrapping the PEC PAD tightly down and around the shaft as shown Secure the PEC PAD strip at the bottom of the shaft with another piece of small tape When finished it should...

Page 5: ...n the galvo block as shown Use the swab to gently rotate the mirror so that the surface is more easily accessible Gently wipe the mirror from the base of the mirror to the end of the mirror Rotate the...

Page 6: ...own Use the swab to gently rotate the mirror so that the surface is more easily accessible Gently wipe the mirror from the base of the mirror to the end of the mirror Rotate the swab and wipe again On...

Page 7: ...ecure the galvo block to the Form 1 frame Reconnect the galvanometer cables and the connect the laser to the main board Then reinstall the Form 1 back panel 9 Plug in Cables Laser Y Galvanometer X Gal...

Reviews: