1. Remove the Back Panel

At the back of the Form 1+, remove two black button-head screws on either 

side using the 2.5mm hex wrench. Lift the cover and remove two silver 

countersunk screws under the cover. Some machines may also have two 

silver countersunk screws along the rear edge on the bottom of the Form 

1+.

Carefully remove the back panel after all screws have been removed.

Properly label the galvanometer cables so they will not be switched when 

plugging them back in.  Unplug the connectors to the Y-Galvanometer, the 

X-Galvanometer, and the laser.  

2. Unplug Laser

Y Galvanometer

X Galvanometer

Laser

Summary of Contents for GALVO

Page 1: ...ace mirrors in the Form 1 can cause scratches and irreversible damage Exercise extreme caution when the mirror is exposed Contact support formlabs com with any questions Formlabs Customer Support can...

Page 2: ...screws under the cover Some machines may also have two silver countersunk screws along the rear edge on the bottom of the Form 1 Carefully remove the back panel after all screws have been removed Prop...

Page 3: ...m the Form 1 frame WARNING DO NOT REMOVE OR TAMPER WITH ANY OTHER SCREWS IN THE GALVO BLOCK ADJUSTING ANY OTHER SCREWS WILL MOVE THE PRINTER OUT OF CALIBRATION Cut a strip about 1cm wide out of PEC PA...

Page 4: ...PAD over to the side Begin wrapping the PEC PAD tightly down and around the shaft as shown Secure the PEC PAD strip at the bottom of the shaft with another piece of small tape When finished it should...

Page 5: ...n the galvo block as shown Use the swab to gently rotate the mirror so that the surface is more easily accessible Gently wipe the mirror from the base of the mirror to the end of the mirror Rotate the...

Page 6: ...own Use the swab to gently rotate the mirror so that the surface is more easily accessible Gently wipe the mirror from the base of the mirror to the end of the mirror Rotate the swab and wipe again On...

Page 7: ...ecure the galvo block to the Form 1 frame Reconnect the galvanometer cables and the connect the laser to the main board Then reinstall the Form 1 back panel 9 Plug in Cables Laser Y Galvanometer X Gal...