DRT-301M User Manual
Shanghai Fengyilong Electronic Co,. Ltd.
www.china-meters.com
Email:[email protected]
021-60975591/ 60975592
Problem
Check
Solution
No communication
Of RS485 wire data
with the meter
Is the meter ID
correct?
Is the
communication
distance too long?
Is there too many
meters connected
to RS485 main wire
Is the RS485 port
connection correct?
Is the connection
Maybe there is a
fault in the inside
circuit.
Check and use the correct the
Meter ID. After the meter finish
the production, the Meter ID
defaults to 0
~
67H ,69H
~
FFH
(Modbus protocol)
or 0000000000000
~
999999999999(DL/T645
protocol) .
Shorten the communication
distance between the reading
equipment outside and the meter.
Keep sure it is not more than
1200m
The equipment to connection with
RS485 main wire is not more than
256pcs
The correct connection is: the A
signal wire of RS485 main wire to
the meter terminal 11,the B signal
wire of RS485 main wire to the
meter terminal 10
Please connect with technical
supporter to replace this meter.
No communication
of the infrared
wireless data with
the meter
Is the meter ID
correct?
Is the
communication
distance too long?
Is the
communication
protocol correct?
Maybe there is a
fault in the inside
circuit.
Check and use the correct the
Meter ID. After the meter finish
the production, the Meter ID
defaults to 0
~
67H ,69H
~
FFH
(Modbus protocol)
or 0000000000000
~
999999999999(DL/T645
protocol)
Shorten the communication
distance between the reading
equipment outside and the meter.
suggest not more than 5m
Please contact the technical
support to get the meter
communication protocol
Please contact the technical
supporter to replace this meter.