In normal working conditions, the data valid LED will come on when the power-up
time has elapsed.
1.15
Zero Calibration
During the commissioning of a new device it is required to perform a Zero
Calibration.
To perform a Zero Calibration of the device, follow the menu Path: Setup Mode
→
Zero Calibration
→
Full Zero Cal, and press Enter to CAL.
Once the Zero Calibration has begun the Data valid light will extinguish, and the
user will be returned to the Zero Calibration Sub-Menu. If the user re-enters the
menu “Wait for FZC” shall be displayed
until the Zero Cal is completed. The Zero
Calibration will take 5 minutes to complete.
Main Menu
Submenu 1
Submenu 2
Setup mode
Zero Calibration
Full Zero Cal
BIAS Current Cal
Zero Ref Time
Zero Frequency
Exit
During the Commissioning of the device, it is required that the user enters the “Setup Mode” menu, which is
password protected. This PassCode will be prompted each time the user enters the Setup Mode Menu.
the Default PassCode is 0000.
NOTE