8
ABOUT THIS SYSTEM
Intended Use
The Fora Comfort TD-3260 2 in 1 Blood Glucose plus Blood Pressure
Monitoring System is indicated for the quantitative measurement of glucose in
fresh whole blood (capillary blood from the finger, the palm, the forearm, the
upper-arm, the calf and the thigh) for self testing by persons with diabetes in
the home or by healthcare professionals in healthcare facilities. The system is
also intended to be used to measure non-invasively the systolic and diastolic
blood pressure and pulse rate of an adult individual, over age 16, at home.
Principle of Measurement
Blood glucose is based on the measurement of electrical current generated by
the reaction of glucose with the reagent of the strip. The monitor measures the
current and displays the corresponding blood glucose level. The strength of the
current produced by the reaction depends on the amount of glucose in the
blood sample.
Blood pressure is measured non-invasively at the wrist based on the
Oscillometric method.
Both functions work separately (one measurement either blood glucose or
blood pressure at a time) to avoid any interference problems.
Major Features
The greatest feature of this system is that you can view results on the PC
through wireless connection using Zigbee Gateway technology. It provides a
simply way to transmit data into your PC.