5/65
General Blood Glucose Meter questions:
Q1:
How is the test strips used with FORA’s blood glucose monitors?
Blood Glucose Only:
Model Name
ACS045
ACS047
ACS044
ACS042
Applicable meter Name
FORA 6
Plus/ Connect/ Smart/
GTel
FORA
Advance Pro/
Comfort Pro
FORA
Prima/Voice/
Mini/GD50
FORA Comfort
Plus/ advance/
plus mini/plus voice
Enzyme Type
GDH-FAD
GDH-FAD
GDH-FAD
GOD
Sample Size
0.5 uL
1.1 uL
0.5 uL
0.5 uL
Reaction Time
5 Seconds
5 Seconds
5 Seconds
5 Seconds
Hematocrit Range
0 ~ 70%
0 ~ 70%
20 ~ 60%
20 ~ 60%
Measuring Range
10 ~ 600 mg/dL
(0.55 ~ 33.3 mmol/L)
10 ~ 600 mg/dL
(0.55 ~ 33.3 mmol/L)
20 ~ 600 mg/dL
(1.1 ~ 33.3 mmol/L)
20 ~ 600 mg/dL
(1.1 ~ 33.3 mmol/L)
Alternative Site Testing
YES AST
No AST
YES AST
YES AST
Applicable Sample
(Whole Blood)
Capillary; Venous;
Arterial
Capillary; Venous
Capillary;
Venous
Capillary
Intended use
General Patient; Dialysis;
Gestational; Neonatal
General
Patient; Dialysis;
Gestational; Neonatal
General Patient; Dialysis;
Gestational
General Patient
Code Type
NO Code Card
NO Code Card
NO Code Card
NO Code Card