Q33: Common Abnormal Results from Control Solution Test.
Item
Outcome of value
Root Cause
Solution
1
The results might be higher.
There is some residue on
the carrier (cap or tip) of the
control solution.
Clean the carrier (cap or
tip of control solution) and
test again. This time use
the second drop of the
control solution.
2
The results might be lower.
The
pollutants
are
absorbed in the control
solution.
Replace
the
control
solution and test again.
3
The results might be higher or
lower.
(1) It is not in the correct
user mode (e.g., not in
CTL
or
QC
Mode
during
the
control
solution test).
(2) You may be using a
different
control
solution.
(3) Control solution may
have expired.
(1) Choose the correct
user mode and test
again.
(2) Use ForaCare control
solution only.
(3) Discard 90 days after
opening.
4
The results might be lower.
Insufficient control solution
is absorbed.
Please test again and
remove the strip only after
hearing the "BEEP" sound
from the meter.
5
The variance might be larger.
The temperature of the
strips
and
meter
is
unstable.
Please test again after the
strips and meter have
been placed in the same
environment
for
30~60
minutes.