20
Step 3
Apply Control Solution.
Shake the control solution vial thoroughly before use. Squeeze out a
drop and wipe it off, then squeeze another drop and place it on the tip
of the vial cap.
Hold the monitor to move the absorbent hole of test strip to touch the
drop. Once the confirmation window fills completely, the monitor will
begin counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
Step 4
Read and Compare the Result.
After counting down to 0, the test result of control solution will appear
on the display. Compare this result with the range printed on the test
strip vial and it should fall within this range. If not, please read the
instructions again and repeat the control solution test.
Summary of Contents for D20 No-Coding
Page 58: ...58 ...