Accessories
Temperature alarm
Before installing alarm
1. Remove supplied 9-volt back-up battery
from packing box.
2. Remove 2 screws from module face and
remove faceplate.
3. Install back-up battery on battery connector.
4. Locate DIP switches on the back of the
faceplate (Fig. 9).
5. Review the factory DIP switch settings
(Fig. 10) and make any changes required to
meet the needs of your specific application.
6. Reinstall faceplate.
Dip Switch
OFF (factory default setting)
ON
1
No alarm delay
45 minute alarm delay
2
Auto reset (alarm stops automatically as
Manual reset of alarm (user must press
soon as temperature returns to set range)
RESET button to stop audible alarm even if
temperature has returned to set range)
3 & 4
Audible alarm on
No audible alarm
3
Enables 5 min alarm “snooze”
Enables 45 min alarm “snooze”
when reset button is pressed (with 4 on)
when reset button is pressed (with 4 off)
Fig. 10
Fig. 11
Installing alarm
1. Mount alarm in desired location with screws (supplied
by others) through back of housing. Do not place
undue strain on probe cable.
Note: Do not modify length of probe wire. Probe will not
measure temperature correctly if wire length is changed.
2. Plug power cord into 110 outlet.
3. Push center tab of short bottle bracket into top of rear left
pilaster (Fig. 11).
Note: Use longer bottle bracket for REF5 refrigerators
using cut-out upper shelf accessory.
4. Fill bottle with glycerin or other liquid to increase accuracy
of readings by simulating the internal temperature of
medications.
5. Insert bottle into bottle bracket.
11
Fig. 9