01441492R00 – REFVAC20/25 LB, PH, PD
7
Fill product bottle
Fig. 4
ATTENTION INSTALLER
§
It is imperative that the top/bottom probe bottle be filled with a solution
that your facility uses to ensure proper operation of unit .
§
Units are equipped with a third-party monitor installation port .
Customers are required to refer to CDC Vaccine and Handling Toolkit
(cdc .gov) for proper placement of the probe within the cabinet to
ensure proper performance .
§
NSF 456 requires the temperature display probe (P4-RTD1 in
REF12, REF20/25, FZR20/25, and REF45 models) to be replaced or
recalibrated . Follett recommends replacement/recalibration 12 months
from the original date of calibration on the Certificate of Traceability
supplied with your equipment at the time of purchase . Recalibration
or purchasing a new calibrated probe is not covered by the original
equipment warranty . To purchase a new Probe with Certificate of
Traceability, please contact our Parts Department @ 1-877-612-5086
or partsales@follettice .com
1. Remove probe and cap from the top bottle and fill with glycerine
supplied with unit (e .g ., 50% glycerine and 50% water)
(Fig. 4) .
Replace cap and reinsert probe sensor .
2. Adhere supplied probe wire clips (x4, packaged with manual) to
inside wall leading to the probe bottle .
3. Route probe wire into clips and close the clips .
Power the Unit
Fig. 5
1. Press lower left corner of front facade door to access power switch
inside
(Fig. 5.1) .
Fig. 6
2. Press power switch to turn unit on (Fig. 6.1) .
3. Connect wiring harness to back-up battery pack .