Operating Manual EMGZ622A/642A
38
Error
Cause
Corrective action
Gain of channel n cannot
be switched
The dig. input used is wrong
parametrized
Set global parameter
dig. input 1...4
to
Switch gain C.n
Limit switches of channel
n do not work
Limit values wrong
parametrized
Set channel parameter
Min. limit / Max.
limit
to an appropriate value
The dig. outputs used are
wrong parametrized
Set global parameters
dig. output 1...4
to
Min. limit n / Max. limit n
Dig. outputs do not work
Wiring error
Check wiring of the dig. outputs (open
collector, ref. to wiring diagram)
C.n Overcurrent
Excitation of channel n detects
overcurrent (short circuit)
Check force sensors and wiring of
channel n
C.n Cable break
Excitation of channel n detects
cable break
Check force sensors and wiring of
channel n
C.n HW error
Hardware of channel n defect
Contact FMS customer service
Subprint of channel n is not
detected
Check if subprints are seated correctly
(ref. to „14.1 Additional Setting
Elements“)
Contact FMS customer service
Subprint missing
contact FMS AG
One or more subprints are
missing or are not detected
Check if subprints are seated correctly
(ref. to „14.1 Additional Setting
Elements“)
Contact FMS customer service
System Error
contact FMS AG
Electronic unit defect
Contact FMS customer service
No message on the
display
Display contrast setting is bad
Set display potentiometer P200 correctly
(ref. to „14.1 Additional Setting
Elements“)
Fuse blown
Replace fuse (ref. to „14.1 Additional
Setting Elements“)
Power supply not correct
Check status LED’s of the power supply
(D111...D115, ref. to „14.1 Additional
Setting Elements“)
Check / correct power supply
Electronic unit defect
Check status LED’s of the power supply
(D111...D115, ref. to „14.1 Additional
Setting Elements“)
Contact FMS customer service
While power failure, no
error message is
provided
Dig. output is parametrized to
Error
Parametrize dig. output to
Ok
and invert
signal input of the following evaluating
device
Electronic unit does not
answer to interface
Interface wrong parametrized
Parametrize interface correct (ref. to „10.
Serial Interface“ ... „13. Interface
DeviceNet“)
commands
Interface wrong connected
Check / correct wiring (ref. to wiring
diagram)
Summary of Contents for EMGZ622A
Page 8: ...Operating Manual EMGZ622A 642A 8 6 Dimensions fig 3 Dimensions E611002e...
Page 26: ...Operating Manual EMGZ622A 642A 26 10 Serial Interface RS232 Optional...
Page 28: ...Operating Manual EMGZ622A 642A 28...
Page 31: ...Operating Manual EMGZ622A 642A 31 13 Interface CAN Bus Optional...
Page 32: ...Operating Manual EMGZ622A 642A 32 14 Interface DeviceNet Optional...