2-6
Getting started - Interfaces
Digital
The yellow channel is used as the:
- Input for digital signals from a trigger or a RPM sensor
- Serial interface for data transmission (RS 232)
- Output for the analog signal (connection for headphone/
oscilloscope)
- Output for stroboscope control (TTL signal)
The permissible range for trigger signals is:
-26V ... 0V (negative) or -5V ... +26V (positive).
Switching threshold positive:
max. 2.5V rising
min.
0.6V
falling
Switching threshold negative:
min. -8V rising
max.
-10V
falling
The input signal must not exceed the switching threshold of the
positive and negative range as otherwise incorrect measurements
can occur.
Temperature
This interface is used for the connection of a thermo couple type K.
Communication
The green channel is used as the:
- Communication interface to the PC. The connection is established
via a network (Ethernet) or directly via a USB cable.
- USB printer port.
- Connection interface for an USB flash drive.
The network is configured in the 'Device Setup' (see 'Ethernet' on
page 2-34).
Attention!
B
A
B
A
B
A
Summary of Contents for VIB 5.310
Page 1: ...Data Collector Vibration Analyzer VIBXPERT II Users Manual ...
Page 9: ...1 8 Empty page ...
Page 51: ...2 38 Empty page ...