Battery
W
Warning
The 9-volt alkaline battery provided with the ULT800 may
explode or leak if recharged, inserted improperly, disposed
of in a fire, or mixed with different battery types. Dispose
of the battery in accordance with any applicable state or
local regulations.
The ULT800 uses a standard 9-volt alkaline battery. The battery has a life
expectancy of approximately 1000 measurements. Replace the battery
yearly, regardless of its condition.
Calibration
W
Warning
Examine the calibration label on the back of the ULT800
prior to each use. Do not use a ULT800 with an expired
calibration label. A ULT800 without a calibration label or
with the anti-tamper case label broken is out of calibration.
A ULT800 that is out of calibration can cause excessive
leakage current exposure to the patient; risk of injury to
the patient could result.
W
Caution
Avoid dropping the ULT800 or allowing other mechanical
abuse that could cause a shift in the ULT800’s calibrated
settings.
The ULT800 requires yearly Fluke factory calibration, which uses
appropriate tools and reference instruments that are traceable to the
National Institute of Standards and Technology (NIST). Factory
calibration provides a calibration sticker on the back of the ULT800 to
verify that the calibration was performed.
To locate a service center, visit the Fluke web site at www.fluke.com, or
contact Fluke at [email protected]. Call from anywhere in the world at
+1-425-446-5500 or call for service in the USA at 1-888-99-FLUKE (1-
888-993-5853.)