Using
the
ULT800
9
TEE TRANSDUCER
LEAKAGE CURRENT TESTER
ULT800
INSTRUCTIONS:
1. Connect the transducer and the adapters to the ULT 800.
2. Press the ON/TEST button and wait for the READY light.
3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.
4. Select LEAKAGE and press the ON/TEST button again-
observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.
ON
READY
PASS
FAIL
LOW BAT
TEST
CONDUCTIVITY
LEAKAGE
Biomedical
600/212 Dual
Conductivity
Electrode
with 600/213
Extenders
ATL
Disinfection
and Storage
Tube
600/210
HP/Agilent
Transducer Adapter
Fluke
ULT800
X
ecy03f.eps
Figure 3. ATL Disinfection/Storage Tube Setup