![Fluke RUSKA 7050 User Manual Download Page 51](http://html1.mh-extra.com/html/fluke/ruska-7050/ruska-7050_user-manual_2310698051.webp)
Remote Operation
Device Messages
5
5-5
:CALibration
:POINts?
Number of calibration points
:VALue?
Nominal
calibration
point
:ZERO <number>
Zero sensor to value
:PRESsure3 Reference
sensor
Same as PRESsure2 above
:PRESsure4PDCR
sensor
Same as PRESsure2 above
:VACuum
Same as PRESsure2 above
:TEMPerature
Gas
Temperature
Same as PRESsure2 above
:TEMPerature2
High
Sensor
Temperature
Same as PRESsure2 above
:TEMPerature3
Low
Sensor
Temperature
Same as PRESsure2 above
:MODE?
return 1 if calibration edit enabled
:MODE <access code>
Request calibration edit
DISP
:ENABle ON|OFF|1|0
Turns Front Panel Display On/Off
TEXT <string>
Displays Message on Front Panel
:BGRaph<number>
Sets Bar Graph Maximum
SENSE
[:PRESSure]
[:RESolution] <number>
Set Pressure Display Resolution
:AUTO<boolean>|ONCE Return to Default Resolution
:MODE?
Returns ABSOLUTE or GAUGE
:MODE ABSolute|GAUGe
Turns On or Off Simulated Absolute
:RANGE
[:UPPer]?
Returns PPI Full Scale Value in Units
:LOWer?
Returns Lowest Calibrated Value
:REFerence
[:HEIGht] <number>
Set Gas Head Height
:MEDium N2|AIR
Set Gas Medium
STATus
:OPERation
[:EVENT]?
Read/Clear Operation Event Register
:CONDition?
Read Operation Condition Register
:ENABle<number>
Set Operation Enable Mask
:QUEStionable
[:EVENT]?
Read/Clear Questionable Event Register
:CONDition?
Read Questionable Condition Register
:ENABle <number>
Set Questionable Enable Mask
:PRESet
Reset
Condition
Flags
SYSTem
:DATE <year>,<month>,<day>
Set System Date
:ERRor?
Returns
<error#,”descr;info”>
Or
0,”No
Error”
:KLOCk ON|OFF|1|0
Lock Keyboard
:TIME
<hour>,<minute>,<second>
Set System Time
:VERSion?
Returns
1991.0
:LANGuage “6000”|”SCPI”|"500" Set Interface Protocol to 6000 or SCPI, or
Druck DPI-500
:PRESet
Reset
System
TEST
:ELECtronic?
Perform
Electronic
Self-Test
STOP
Abort
Pneumatic
Self-Test
UNIT
:DEFine<n> <name>,<number>
Define a Unit
:LENGth MM|IN
Set Length Units for Head Height
[:PRESsure] <unit name>
Set Pressure Units. Valid unit names are
KPA,
BAR,
PSI,
KG/CM2,
MMHG0C,
CMHG0C,
INHG0C,
INHG60F
CMH204C,
INH2020C,
INH2025C,
%FS
FT, M, KNOT, KM/HR, MPA, PA
Summary of Contents for RUSKA 7050
Page 3: ...7050 Change Language Hold key for 5 seconds mode enter...
Page 4: ......
Page 10: ...RUSKA 7050 7050i 7050LP Users Manual vi...
Page 12: ...RUSKA 7050 7050i 7050LP Users Manual viii 7 1 Packing the PPI 7 3...
Page 18: ...RUSKA 7050 7050i 7050LP Users Manual 1 6...
Page 32: ...RUSKA 7050 7050i 7050LP Users Manual 3 4...
Page 60: ...RUSKA 7050 7050i 7050LP Users Manual 5 14...