Model 787
Calibration Manual
22
DC Millivolts Measurement Test
1.
Put the calibrator in Standby (
STBY
) mode.
2.
Put the UUT rotary switch in the
m
L
(dc Millivolts)
position.
3.
Connect the calibrator to the
COM
and
z
terminals on the UUT as shown in
Figure 9.
4.
Apply the values from the calibrator shown in Table 8 and compare the readings on
the UUT to the acceptable readings shown.
Table 8. DC mV Test
787 Range
Calibrator DC Voltage
Minimum Reading
Maximum Reading
No Range Switching
100 mV
99.8 mV
100.2 mV
No Range Switching
300 mV
299.6 mV
300.4 mV
OFF
A
mA
COM
V
1000V
30mA
FUSED
0.44A
(1A /30 sec)
FUSED
mA
mA
A
mA
OUTPUT 0-24mA
SOURCE SIMULATE
+
+
% STEP
COARSE
FINE
mV
V
V
OUTPUT
CAT
787
PROCESSMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
UUT
5500A
POWER
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
SCOPE
BOOST
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
÷
OUT
TRIG
5500A CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
LT0005F.EPS
Figure 9. DC mV Measurement Test Connections
Summary of Contents for ProcessMeter 787
Page 2: ......
Page 4: ...Model 787 Calibration Manual ii ...
Page 6: ...Model 787 Calibration Manual iv ...