Fluke P5523 User Manual Download Page 15

 

Liquid to Gas Separator 

 

Ancillary Equipment 

 

 

3.

 

When the fluid has been withdrawn from the separator, the connection hose can be 

disconnected from the deadweight tester.  

Periodically, the right-hand reservoir body should be inspected and cleaned, to remove 

any fluid that may be present.  

Ancillary Equipment 

Angle Adapter, P5543 

To calibrate gauges with the pressure connection on the rear (e.g. panel-mount gauges) in 

their correct position, an angle adapter should be used. The angle adapter fits directly 

onto the test station, converting it through 90 degrees, allowing the standard adapters to 

be used. 

 

gle09.bmp 

Figure 4. Angle Adapter 

Pointer Remover/Punch, P5551 

To remove and refit the pointer of a pressure gauge, use the Pointer Remover/Punch. This 

tool has a spring-loaded plunger to quickly and consistently refit the pointer. 

 

gle10.bmp 

Figure 5. Pointer Remover/Punch 

Summary of Contents for P5523

Page 1: ...2010 2010 Fluke Corporation All rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies P5523 Liquid to Gas Se...

Page 2: ...vice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no...

Page 3: ...pressed Liquid Gas 2 Personal Protective Equipment 2 Symbols Used in this Manual 2 Preparations 3 Connect the Gauge 4 Connect the Test Port Insert 4 Priming 5 Operation 6 Disconnecting the Unit from t...

Page 4: ...P5523 Users Manual ii...

Page 5: ...iii List of Tables Table Title Page 1 Symbols 2 2 Test Port Insert Parts List 5...

Page 6: ...P5523 Users Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 Hydraulic Pneumatic Circuit Schematic 1 2 Connecting the Gauge 4 3 Test Port Insert 5 4 Angle Adapter 7 5 Pointer Remover Punch 7...

Page 8: ...P5523 Users Manual vi...

Page 9: ...ibrators and Transducers An external high pressure gas source is applied through a series of liquid traps to the hydraulic system of the deadweight tester Typically dry Nitrogen from a commercially av...

Page 10: ...ent operation and maintenance to ensure safety and health and protection of property Compressed Liquid Gas Use of compressed liquids and gasses can create an environment of propelled foreign matter Pr...

Page 11: ...ensure that the supply is clean and dry W Caution The external pressure supply must be regulated to either the maximum range of the P5523 3000 psi 210 bar or 10 above the maximum pressure required wh...

Page 12: ...to face forward hold the gauge adapter and turn the instrument COUNTER CLOCKWISE so that it faces forward 4 Hold the instrument steady while turning the gauge adapter COUNTER CLOCKWISE until it pulls...

Page 13: ...tion Ancillary Equipment Priming 1 Unscrew and remove reservoir cap from left hand reservoir Note the position of the groove inside the reservoir body this is the Fluid Datum Level 2 Using the priming...

Page 14: ...re to rise or fall which can cause instrument readings to change as the fluid in the system expands or contracts thus increasing or decreasing the pressure 6 Compare the reading of the gauge under tes...

Page 15: ...rate gauges with the pressure connection on the rear e g panel mount gauges in their correct position an angle adapter should be used The angle adapter fits directly onto the test station converting i...

Page 16: ...P5523 Users Manual 8...

Reviews: