P5523
Users Manual
4
1
2
3
4
gle11.eps
Figure 2. Connecting the Gauge
Connect the Gauge
Refer to Figure 2 for steps 1-4.
1.
Screw the appropriate gauge adapter fully on to the instrument to be tested.
2.
Screw assembly down COUNTER-CLOCKWISE onto test port.
Note
Hand-tight is sufficient; ensure that the bottom face contacts the O-ring on
the test port.
3.
To adjust the position to face forward, hold the gauge adapter and turn the instrument
COUNTER-CLOCKWISE, so that it faces forward.
4.
Hold the instrument steady, while turning the gauge adapter
COUNTER-CLOCKWISE until it pulls down onto the O-ring.
Connect the Test Port Insert
For devices with 1/8 BSP or NPT mounting threads, the diameter of the thread is very
close to the effective sealing diameter of the O-ring fitted to the test port.
This can make it difficult to achieve a good seal. When mounting these devices, use the
test port insert (stored in the spare seals container) as shown in Figure 3.