Liquid to Gas Separator
Priming
5
1
2
3
4
gle012.eps
Figure 3. Test Port Insert
Table 2. Test Port Insert - Parts List
Item Description Part
Number
1 O
Ring
3865142
2
Test Port Insert
3919892
3 O
Ring
3883397
4
Test Port
—
To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle
Adapter. (Refer to section, Ancillary Equipment.)
Priming
1.
Unscrew and remove reservoir cap from left-hand reservoir. Note the position of the
groove inside the reservoir body — this is the Fluid Datum Level.
2.
Using the priming pump on the deadweight tester, slowly drive fluid from the
deadweight tester through the connection pipe, into the left-hand reservoir, until it is
just below the Fluid Datum Level.
3.
Allow a few moments for the system to stabilize, and then bring the fluid up to the
Datum Level by slowly turning the screw press in (clockwise).