Do the touch targets seem to be off?
Try touching around the target to verify if the touch target calibration is off (this is
rare). If you suspect the touch target is off, select
Setup | Display
, press
Touch
Screen Controller…
, then
Calibrate
. You will be prompted to touch the display
at certain points to perform the calibration.
Are you connected to the network?
!
Link light should be on solid green or amber if Link exists.
!
Some activity on Transmit LED and/or Utilization LEDs should be seen.
!
Go to
Setup | Ethernet
screen. An active interface should be indicated.
!
Confirm that Cable Test passed. Try a different cable if necessary.
Does Cable Test pass?
A network connection can not be made if cable test does not pass. Look at the
Front Page
screen
Cable Test
button and see if there is a red X or green check
mark in the button. A red X could indicate a faulty cable. If there is a red X, press
the
Cable Test
button and go to the
Cable Test | Overview
and
Twisted-Pair
Detail
screens to test the cable.
Does the battery charge state appear erratic or inconsistent, or is the battery not
consistently holding a normal charge?
Discharge the battery completely, then charge the battery for at least 5 hours.
Does the touch panel display (or Windows) not respond to input?
Press the
On/Off
button to place the analyzer in standby. Press the
On/Off
button
again to resume from standby.
If this does not correct the problem, try using a keyboard and/or mouse to
shutdown and restart Windows.
If this does not correct the problem, try pressing the
Reset
button (see Figure 28)
to reset Windows.
Note
A Windows Reset will not erase information collected on the data
acquisition card.