 3.

  Do the OTDR Test

W

Caution

Do not connect PC or UPC connectors to the HDR OTDR port. Doing so can 

damage the port’s APC endface, and the large reflection at the OTDR port 

can cause unreliable test results.

3-1 

Inspect all connectors and clean as necessary.

3-2 

Connect the tester to the PON link (Figure 3). 

3-3 

Tap 

TEST

 or press 

.

SC-APC

BV02.EPS

Figure 3. 

Typical Connection to an Out-of-Service PON Link

 4.

 

Examine the Results

The tester shows the OTDR results in three formats: 

BV26.EPS

Figure 4. Examples of PON OTDR Results Screens

3

Short patch 

cord and 

adapter

Connector from ONT

(Optical Network Terminal)

Feeder link

Optical Line 

Terminal at 

Central Office

ONT

OLT

Splitter

4

Summary of Contents for OptiFiber PRO OTDR

Page 1: ...on see the latest versions of the Versiv Series Users Manual and the Versiv Series Technical Reference Handbook provided on the Fluke Networks website About LinkWare Live With a Fluke Networks web account and a wired or wireless network connection you can use the LinkWare Live web application to store and manage projects in the cloud For more information go to www linkwarelive com signin or see th...

Page 2: ...he WEEE Directive Annex I this product is classed as category 9 Monitoring and Control Instrumentation product Do not dispose of this product as unsorted municipal waste To return unwanted products contact the manufacturer s web site shown on the product or your local sales office or distributor P Conformite Europeene Conforms to the requirements of the European Union and the European Free Trade A...

Page 3: ...ed as recharged twice a week Heavy use is defined as discharged to cutoff and recharged daily Disconnect the battery charger and move the Product or battery to a cool non flammable location if the rechargeable battery becomes hot 50 C 122 F during the charge period The battery door must be closed and locked before you operate the Product Repair the Product before use if the battery leaks Recharge ...

Page 4: ...port turns on when the port emits an optical signal C Visual fault locator port and protective cap The LED in front of the port turns on when the port emits an optical signal D Button that controls the VFL E Micro USB port This USB port lets you connect the tester to a PC so you can upload test results to the PC and install software updates in the tester F Type A USB port This USB host port lets y...

Page 5: ... when the battery charges and green when the battery is fully charged The LED is yellow if the battery will not charge L RJ45 connector Lets you connect to a network for access to Fluke Networks cloud services M Decal with laser safety information WWarning Do not look directly into optical connectors Some sources emit invisible radiation that can cause permanent damage to your eyes ...

Page 6: ...The Manual PON OTDR test includes Splitter Setup settings that let you enter Splitter Ratios To locate and identify splitters tap DISCOVER on the SPLITTER SETUP screen Launch Compensation This function removes the effects of launch and tail cords from the test results To use it select On then tap SET COMP on the home screen to select a compensation method Macrobend Detection The default of 0 50 dB...

Page 7: ...to LinkWare Live For more information see the Fluke Networks website or the Users Manual The Auto PON OTDR test automatically identifies splitters To enter settings for splitters select a Manual PON OTDR test Set up a project Change the cable ID and Auto Save setting ...

Page 8: ... clean as necessary 3 2 Connect the tester to the PON link Figure 3 3 3 Tap TEST or press SC APC BV02 EPS Figure 3 Typical Connection to an Out of Service PON Link 4 Examine the Results The tester shows the OTDR results in three formats BV26 EPS Figure 4 Examples of PON OTDR Results Screens 3 Short patch cord and adapter Connector from ONT Optical Network Terminal Feeder link Optical Line Terminal...

Page 9: ... Fluke Networks 6920 Seaway Boulevard MS 143F Everett WA 98203 USA Fluke Networks operates in more than 50 countries worldwide For more contact information visit our website General Specifications Battery Type Lithium ion Charging Temperature 0 C to 45 C 32 F to 113 F Power Adapter Input 100 to 240 VAC 10 50 60Hz Output 15 VDC 2 A maximum Class II Versiv 2 wireless radio Frequency ranges 2 4 GHz b...

Page 10: ...s behalf To obtain service during the warranty period contact your nearest Fluke Networks authorized service center to obtain return authorization information then send your defective product to that Service Center with a description of the problem For a list of authorized resellers visit www flukenetworks com wheretobuy THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A P...

Reviews: