Output
The rightmost terminals are used to connect the scanner to the Input 1 measure-
ment input of the 1590 Thermometer.
Channel Inputs
Each of the ten channels has four terminal posts with which a probe is con-
nected. The red and black current terminals (“C1" and ”C2") supply current to
the probe. The red and black potential terminals (“P1" and ”P1") are used to
sense the voltage on the probe. Guard connector
The guard or shield of the probe cable can be connected to the “Guard” termi-
nal. If the guard is used, the Output guard terminal of the scanner should be
connected to the Input 1 guard terminal of the 1590. All of the guard terminals
of the inputs are wired to the output guard terminal; they are not switched. Us-
ing a guard may help to reduce the effects of interference in electrically noisy
environments.
Channel Select
The ten front panel buttons allow you to select the channel for measurement.
The channel can also be selected by the 1590 Thermometer.
Channel indicator
The channel indicator is shows which of the scanner channels, if any, is se-
lected for measurement.
Control Input (Rear Panel)
The control input is located on the rear panel of the scanner (see Figure 20).
The control cable provided with the scanner connects the control input of the
scanner to the control output of 1590. The control interface provides the control
signals and the power necessary to operate the scanner. The scanner will not
operate without being properly connected to the 1590.
11.5
Scanner Operation
This section explains operation of the 1590 Thermometer with the 2590 Scan-
ner using the front panel buttons and soft-keys.
129
11 2590 Scanner (optional)
Scanner Operation