9
Calibration
This chapter explains the step-by-step procedure for calibrating and verifying
the 1575/1590 Thermometer to ensure that the instrument meets its published
specifications. The calibration should be performed at least once per year. In
line with normal prudent metrology practices, Hart recommends a short-cycle
calibration interval of six months for new units during the first year to ensure
that all components are as stable as expected.
9.1
Setup
The following standard resistors are recommended for calibrating and verifying
the resistance accuracy of the thermometer:
The 1575/1590 must be turned on and allowed to warm up for at least one hour
before proceeding with the calibration. The resistor oven should also be on and
warmed up for at least one hour. (See Section 7.6.5.2 to operate the resistor
oven.)
109
9 Calibration
Setup
Standard Resistor
Uncertainty
1
Ω
10 ppm
10
Ω
5 ppm
100
Ω
1.5 ppm
10 k
Ω
2.5 ppm
Table 15
Standard Resistors Used for
Calibration of the Reference Resistors
Standard Resistors
Ratio Uncertainty
25
Ω/
100
Ω
0.25 ppm
100
Ω
/100
Ω
0.25 ppm
400
Ω
/100
Ω
0.25 ppm
Table 16
Standard Resistors Used for
Calibration of the Ratio Measurement