Fluke DSP-FOM Service Information Sheet Download Page 6

DSP-FOM
Fiber Optic Meter

6

Step 5:  1550 nm Fine Calibration
Adjustment

5-1. Set the switch on the FOM to the 1550 nm

position.

5-2. Set the 1550 nm light source to -10 dBm

±

0.05 dBm. Use the reference power meter to

measure the source output power level.
Record this level as Pin for use in step 5-4.

5-3. Connect the 1550 nm light source to the

FOM.

5-4. Use Pin from step 5-2, Vref from step

2-5, and the transfer function below to
calculate Vo:

 

( )

(

)

9R

= 

 × −

+

V

53 dBm

P

3 dBm

ref

in

5-5. Measure between DBout at U2 pin 7 and the

analog ground reference at U1 Pin 4. Adjust
R3 for Vo (as calculated in step 5-4)

±

0.0005 V.

Step 6:  Verifying Calibration

6-1.

 

Follow the procedures given in the earlier
section "Performance Tests".

6-2.

 

Apply insulating varnish, such as red GLPT,
to R1, R2, R3, and R10.

6-3.

 

Reassemble the FOM as described in the
earlier section "Disassembling and
Reassembling the FOM".

Parts and Accessories

Replacement parts (Table 2) and accessories
(Table 3) are available from Fluke.

Table 2. Fluke Replacement Parts for the DSP-FOM

Description

Fluke Part

Number

DSP-FOM Plastic Case Assembly

616034

Battery Door

200474

PD1: Germanium Photodiode, 1 mm

602778

Table 3. Accessories for the DSP-FOM

Description

Fluke Part

Number

ST to ST Multimode Patch Cable, 1m

617453

ST to ST Cable Adapter

602810

Carrying Case

616091

Summary of Contents for DSP-FOM

Page 1: ...ents Parts and accessories lists Schematic Diagram showing calibration measurement and adjustment points For operating instructions refer to the DSP FOM DSP FTK Fiber Optic Accessory Instruction Sheet...

Page 2: ...the selected wavelength and a full scale voltage reference for determining the battery status Switch Positions and I O Slide switch SW1 is a 2 pole switch One pole pin 5 applies the battery voltage t...

Page 3: ...2 required 2 m UTP patch cable supplied with the CableMeter test tool Performance Tests Use the performance tests to confirm that the FOM is working properly If the FOM fails any of these tests clean...

Page 4: ...n While servicing the FOM always follow guidelines for preventing electrostatic discharge ESD Otherwise ESD can damage sensitive components causing immediate or delayed failure of the FOM A Phillips h...

Page 5: ...tput to the reference power meter Set the attenuator output to 45 dBm 0 1 dBm Record the value shown on the reference power meter as Pin for step 3 3 3 2 Connect the attenuator output to the FOM 3 3 U...

Page 6: ...djust R3 for Vo as calculated in step 5 4 0 0005 V Step 6 Verifying Calibration 6 1 Follow the procedures given in the earlier section Performance Tests 6 2 Apply insulating varnish such as red GLPT t...

Page 7: ...Service Information Sheet for Serial Numbers 79370000 Fiber Optic Meter Schematic for SN 79370000 7 FPM2 00 0300 DB out V ref V det Dt11f eps Figure 1 Fiber Optic Meter Schematic for SN 79370000...

Page 8: ...DSP FOM Fiber Optic Meter 8 FPM2 00 0001 Pin 7 Pin 5 Pin 1 Pin 7 Dt14f eps Figure 2 Calibration Measurement and Adjustment Points for SN 79370000...

Reviews: