Fluke DSP-FOM Service Information Sheet Download Page 3

Service Information Sheet for Serial Numbers < 79370000

Cleaning the Optical Connector

3

between the reference current and the signal
current; therefore, the output is a linear function of
the optical power received by the photo detector. A
typical output at U1, pin 7 is 23 mV per dB of
optical power, plus a 69 mV offset

The reference current adjustment potentiometers
R1, R2, and R3 control the zero offset for the three
calibrated wavelengths. Potentiometer R10 trims
the final output slope to 10 mV per dB. The output
of U1B is buffered by U2B. R7 provides
temperature compensation to null some of the
thermal drift of Q1A and Q1B.

Cleaning the Optical Connector

Most problems with optical power meters result
from contaminated connectors. Therefore, always
clean the connector before troubleshooting or
calibration.

To clean the FOM ST connector, wipe it gently
with an optical-grade tissue or swab dampened with
optical-grade alcohol. To remove loose dirt and
dust from the connector, use filtered, compressed
air.

Always cover the connector with a dust cap when
the unit is not in use.

Required Equipment

The following equipment is required for servicing
the FOM:

 

0-10 V power supply

 

Fluke 187 digital multimeter or equivalent

 

Optical power meter calibrated at 850 nm,
1300 nm, and 1550 nm traceable to NIST
Standards

 

850 nm laser source; Po = -10 dBm

 

1300 nm laser source; Po = -10 dBm

 

1550 nm laser source: Po = -10 dBm

 

Variable MM optical attenuator

 

Variable SM optical attenuator

 

Fluke DSP CableMeter test tool

 

1 m fiber optic patch cables, 2 required

 

2 m UTP patch cable (supplied with the
CableMeter test tool)

Performance Tests

Use the performance tests to confirm that the FOM
is working properly. If the FOM fails any of these
tests, clean the fiber connectors, verify that the
connections are solid, then retest the FOM. If the
FOM still fails a test, it needs calibration
adjustments or repair.

DSP-FOM Detection

1.

 

Turn the CableMeter test tool knob to 

SETUP

.

Use the arrow keys to select 

TEST

STANDARD, CABLE TYPE

; then press

E

. Use 

U D

 (and Page Up/Page

Down softkeys if necessary) to select the
applicable standard for the DSP-FOM; then
press 

E

. Use the arrow keys to select

Multimode

 as the cable type; then press

E

.

2.

 

With the FOM not connected to the CableMeter
test tool, turn the test tool knob to 

SINGLE

TEST

; then press 

E

. Verify that the

test tool displays the following message: 

WARNING THE DSP-FOM IS NOT 

CONNECTED

. Press 

E

to continue.

3.

 

With the FOM switch in the 

OFF

 position, use

the 2-meter UTP patch cable to connect the
FOM to the CableMeter test tool. Press the test
tool 

T

 key. Verify that the test tool displays

the following message: 

WARNING

  

THE

 

DSP-FOM IS TURNED OFF

. Press

E

to continue.

DSP-FOM -10 dBm Checkout

1.

 

Turn on the FOM and the three light sources.
Verify that the FOM power LED is on. Allow
the FOM and the sources to warm up for 10
minutes.

 

Perform the remaining steps using the light sources
in the following order:  (1) 850 nm, (2) 1550 nm,
and (3) 1300 nm.

2.

 

Use a fiber optic patch cable to connect the
light source to the reference power meter.

3.

 

Verify that the reference power meter is
configured to measure the wavelength of the
light source. Adjust the output of the source for
a reading of -10.00 dBm 

±

0.05 dBm. Record

the power reading for later use.

Summary of Contents for DSP-FOM

Page 1: ...ents Parts and accessories lists Schematic Diagram showing calibration measurement and adjustment points For operating instructions refer to the DSP FOM DSP FTK Fiber Optic Accessory Instruction Sheet...

Page 2: ...the selected wavelength and a full scale voltage reference for determining the battery status Switch Positions and I O Slide switch SW1 is a 2 pole switch One pole pin 5 applies the battery voltage t...

Page 3: ...2 required 2 m UTP patch cable supplied with the CableMeter test tool Performance Tests Use the performance tests to confirm that the FOM is working properly If the FOM fails any of these tests clean...

Page 4: ...n While servicing the FOM always follow guidelines for preventing electrostatic discharge ESD Otherwise ESD can damage sensitive components causing immediate or delayed failure of the FOM A Phillips h...

Page 5: ...tput to the reference power meter Set the attenuator output to 45 dBm 0 1 dBm Record the value shown on the reference power meter as Pin for step 3 3 3 2 Connect the attenuator output to the FOM 3 3 U...

Page 6: ...djust R3 for Vo as calculated in step 5 4 0 0005 V Step 6 Verifying Calibration 6 1 Follow the procedures given in the earlier section Performance Tests 6 2 Apply insulating varnish such as red GLPT t...

Page 7: ...Service Information Sheet for Serial Numbers 79370000 Fiber Optic Meter Schematic for SN 79370000 7 FPM2 00 0300 DB out V ref V det Dt11f eps Figure 1 Fiber Optic Meter Schematic for SN 79370000...

Page 8: ...DSP FOM Fiber Optic Meter 8 FPM2 00 0001 Pin 7 Pin 5 Pin 1 Pin 7 Dt14f eps Figure 2 Calibration Measurement and Adjustment Points for SN 79370000...

Reviews: