Fluke CPS-100M Instructions Manual Download Page 3

Contamination Prevention System 

High-Pressure Contamination Prevention System

 

 

Symbols

   

 

 

Symbols 

The symbols shown in Table 1 are found in these instructions or on the Product. 

Table 1. Symbols 

 

 
 
 
 
 
 
 

 

Box Contents 

The Product is either of these kits: 
CPS-40M: 

  CPS-40M-HC40 Test Station 

  Pressure Hose, JIC #4 (AN4) fittings 

  Pressure Hose Adapter Fittings, JIC #4 (AN4): 

1/4 NPT; 1/4 BSP; 7/16-20 (x2) 

  Vacuum Hose, 3/8 inch polyethylene 

  Vacuum Hose Adapter Fittings, 3/8 tube: 

1/4 NPT; 1/4 BSP; 7/16-20 (x2) 

  PK-ADPTR-HC40 Adapter Set 

  CPS Cable 

  Instructions 

CPS-100M: 

  CPS-100M Test System  

  High-Pressure Hose, DH500 Fittings (cone 

and threaded connection compatible with 
Autoclave F250C and HIP HF4) 

  CPS Cable 

  Instructions 

 

 

Symbol 

Definition 

W

 

WARNING, RISK OF DANGER. 

 

Consult user documentation. 

)

 

Certified by CSA Group to North American safety standards. 

P

 

Conforms to European Union directives. 

 

Conforms to relevant Australian Safety and EMC standards. 

 

This product complies with WEEE Directive marking requirements. The affixed label indicates that 
you must not discard this electrical/electronic product in domestic household waste. Product 
Category: With reference to the equipment types in the WEEE directive Annex I, this product is 
classed as category 9 “Monitoring and Control Instrumentation” product. Do not dispose of this 
product as unsorted municipal waste. 

Summary of Contents for CPS-100M

Page 1: ...mum working pressure of 44 MPa 6400 psi CPS 100M has a maximum working pressure of 110 MPa 16 000 psi When the CPS is enabled the pressure controller controls the CPS valves to decrease the test syste...

Page 2: ...at are dangerous to the user A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test WWarning To prevent personal injury Read all safety informa...

Page 3: ...Pressure Hose DH500 Fittings cone and threaded connection compatible with Autoclave F250C and HIP HF4 CPS Cable Instructions Symbol Definition W WARNING RISK OF DANGER Consult user documentation Certi...

Page 4: ...ures are shown in Table 2 The CPS 100M features are shown in Table 3 Table 2 CPS 40M 5 3 2 1 6 10 9 8 7 4 ifc01 eps Number Description Stand Pressure Hose Tee D Vacuum Hose E TEST port connector F LED...

Page 5: ...ystem High Pressure Contamination Prevention System Features 5 Table 3 CPS 100M 4 5 8 7 6 3 1 2 ifc02 eps Number Description Stand High pressure Hose DUT Port D LED Indicator E Body F Drain Tube G Fil...

Page 6: ...or SAE to the EXHAUST port of the pressure controller and tighten with a wrench 4 Place the CPS in desired location noting that test gauges will likely be mounted on top of it 5 Plan the cuts on the...

Page 7: ...creates an unsafe condition do not use thread sealing tape or any other sealing method on the DUT or the adapters The Gauge Adapter sealing system can be hand tight sealed up to 44 MPa 6400 psi Wrench...

Page 8: ...nut counter clockwise 1 4 turn 2 Position the DUT to face the desired direction and tighten the gland nut clockwise re seal Operation CPS Tasks Tasks are accessed by the Setup Tasks menu path on the...

Page 9: ...mination Prevention System Purge 9 Ifc06 eps Figure 2 Cutaway of the CPS 40M HC40 during a Normal Downward Pressure Excursion or Purge Task ifc15 eps Figure 3 Cutaway of the CPS 100M Purge Task during...

Page 10: ...being able to vent pressure Once disabled in the firmware the CPS can be disconnected both electrically and pneumatically Do not leave the CPS physically connected to the controller if it has been dis...

Page 11: ...m Maintenance 11 1 Unscrew the filter cover counter clockwise to release the coalescing filter D and O ring see Table 4 E See Figure 3 3 4 5 ifc05 eps Figure 3 2 Unscrew filter housing F counter clock...

Page 12: ...lter housing F to remove any contaminant particles or residue 5 Fit the O ring mesh filter and spring to housing and screw clockwise into the main body 6 Fit coalescing filter and the O ring to the fi...

Page 13: ...manship for one year from the date of purchase This warranty does not cover fuses disposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operati...