CPS-40M-HC40
Contamination Prevention System
CPS-100M
High-Pressure Contamination Prevention System
Instructions
Introduction
The Fluke CPS-40M-HC40 (CPS-40M) Contamination Prevention System and CPS-100M High-Pressure
Contamination Prevention System (Product) are accessories that protect pressure controllers from
contamination by the Device Under Test (DUT).
•
CPS-40M-HC40 (CPS-40M) has a maximum-working pressure of 44 MPa (6400 psi).
•
CPS-100M has a maximum-working pressure of 110 MPa (16 000 psi).
When the CPS is enabled, the pressure controller controls the CPS valves to decrease the test system
pressure, purge contaminants, and completely vent the test system.
The Product is connected to the pressure controller by a high pressure hose and a control cable. There is an
optional vacuum pump connection (CPS-40M only). All upward pressure changes are accomplished by the
pressure controller. Rough downward changes in pressure are accomplished through the Product's vent
valve. This results in most of the gas being exhausted out of the Product instead of going back to the
controller. Contaminants from the DUT are filtered and liquids drop into a sump. The sump is automatically
purged into a waste bottle when the system is vented. The product includes both mesh and coalescing filters
that are replaceable.