9132 Portable IR Calibrator
User’s Guide
DO NOT
use this unit for any application other than calibration work.
DO NOT
use this unit in environments other than those listed in the user’s guide.
DO NOT
operate near flammable materials.
Use of this instrument at
HIGH TEMPERATURES
for extended periods of time
requires caution.
Completely unattended high temperature operation is not recommended for safety
reasons.
Before initial use, after transport, and anytime the dry-well has not been energized for
more than 10 days, the calibrator must be energized for a dry-out period of 1 to 2 hours
before it can be assumed to meet all of the safety requirements of the IEC1010-1.
The instrument can generate extreme temperatures. Precautions must be taken to prevent
personal injury or damage to objects.
Use only a grounded AC mains supply of the appropriate voltage to power the
instrument. Refer to Section 3.1, Specifications for power details.
The instrument is equipped with operator accessible system fuses. If a fuse blows, it may
be due to a power surge or failure of a component. Replace the fuse once. If the fuse
blows a second time, it is likely caused by failure of a component part. If this occurs,
contact an Authorized Hart Scientific Service Center (see Section ). Always replace the
fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a
higher current rating.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
1.3.2 CAUTIONS
DO NOT plug the unit into 230V if the heater switches and fuse holder read 115V. This
action will cause the fuses to blow and may damage the instrument.
Components and heater lifetime can be shortened by continuous high temperature
operation.
DO NOT use fluids to clean the target surface.
DO NOT change the values of the calibration constants from the factory set values. The
correct setting of these parameters is important to the safety and proper operation of the
calibrator.
DO use a ground fault interrupt device.
Operate the instrument in room temperatures between 5 and 50°C. (41–122° F). Allow
sufficient air circulation by leaving at least 6 inches of space between the instrument and
nearby objects.
The instrument is a precision instrument. Although is has been designed for optimum
durability and trouble free operation, it must be handled with care. Always carry the unit
in an upright position. The convenient fold-up handle allows one hand carrying. The
instrument should not be operated in excessively wet, oily, dusty, or dirty environments.
Do not operate near flammable materials.
Before initial use, after transport, and anytime the dry-well has not been energized
for a “dry-out” period of 1-2 hours before it can be assumed to meet all of the safety
requirements of the IEC 1010-1.
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
1.888.610.7664
Fluke
-
Direct
.com