9132 Portable IR Calibrator
User’s Guide
22
4.1
Proportional band setting
To change the proportional band press “UP” or “DOWN”.
10.0
New proportional band setting
To accept the new setting press “SET”. Press “EXIT” to continue without storing the new value.
S
Accept the new proportional band setting
6.10 Controller Configuration
The controller has a number of configuration and operating options and calibration parameters,
which are programmable via the front panel. These are accessed from the secondary menu after
the proportional band function by pressing “SET”. Pressing “SET” again enters the first of three
sets by configuration parameters – calibration parameters, operating parameters and serial inter-
face parameters. The menus are selected using the “UP” and “DOWN” keys and then pressing
“SET”.
6.11 Operating Parameters
The operating parameters menu is indicated by,
par
Operating parameters menu
The operating parameters menu contains the High Limit parameter. The High Limit parameter
adjusts the upper set-point temperature. The factory default and maximum are set to 126 °C.
For safety, a user can adjust the High Limit parameter down so the maximum temperature set-
point is restricted.
Press “SET” to enable adjustment of the High Limit parameter.
HL
Flashes
HL
and then displays the setting
H=126
Current HL setting
Adjust the HL parameter using “UP” or “DOWN”.
H=90
New High Limit setting
Press “SET” to accept the new High Limit parameter.
6.12 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErIAL
Serial RS-232 interface parameters menu
Press “UP” to enter the menu. The serial interface parameters menu contains parameters, which
determine the operation of the serial interface. The parameters in the menu are – baud rate,
sample period, duplex mode, and linefeed.