Dry-well Calibrator
Test Probe Calibration
23
Table 3. Controller Communications Commands (cont.)
Test Probe Calibration
For optimum accuracy and stability, allow the Product to warm up for 10 minutes after power-up and
then allow adequate stabilization time after reaching the set-point temperature. After completing
operation of the calibrator, allow the well to cool by setting the temperature to 25 °C for one-half hour
before turning off the power.
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set serial duplex mode:
du[plex]=f[ull]/
h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu
Read R0 calibration parameter
r[0]
r0: 999.999
r0: 100.7
Set R0 calibration parameter to
n
r[0]=n
r=100.7
95.0 to 105.0
Read ALPHA calibration parameter
al[pha]
al
al: 9.99999999
al: 0.003865
Set ALPHA calibration parameter to
n
al[pha]=n
al=0.003865
0.002 to 0.006
Read DELTA calibration parameter
de[lta]
de
de: 1.50
Set DELTA calibration parameter
de[lta]=n
de=1.37
de: 9.99999
0.0–3.0v
Miscellaneous Other Commands
Read firmware version number
*ver[sion]
*ver
ver.9999x,9.99
ver.9102S,1.10
Read structure of all commands
h[elp]
h
list of commands
Read ALL operating parameters
all
all
list of parameters
Read set-point
*sr
*sr
999.999 ohms
100.123 ohms
Legend:
[] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by
a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.