Manual Supplement
810 Getting Started
2
3/10
On page 12, replace the
Sensor Setup
and the
Compatible Sensor
paragraphs with the following:
Sensor Setup
The Tester includes a triaxial Sensor with TEDS technology (Transducer Electronic Data Sheets). With this
technology, your Tester can identify and automatically read the Sensor configuration. This technology provides:
‚
Improved results from detailed calibration information
‚
Reduced configuration time without manual data entry
‚
Sensor calibration tracking with last calibration date stored electronically
Compatible Sensors
It is highly recommended to use a Fluke Triaxial Sensor with the Tester. Using a Sensor other than a Fluke
Triaxial Sensor will result in misleading diagnoses. The Tester is compatible with single axis Sensors.
On page 13, add the following bullet under the Caution:
‚
Always disconnect the Sensor cable from the Tester when not in use.
On page 14, delete steps 2, 3 and 4 and add the following Warning:
V
Warning
To avoid injury when attaching reflective tape to the machine, stop the rotating machine. Attach a piece
of reflective tape onto the shaft or other rotating part of the machine. Restart the machine and wait until
it reaches its normal operating conditions.