Fluke 80i-110s Instructions Manual Download Page 8

Battery

Type ......................................... 9 volt, IEC 6LR61
Consumption ............................ 8.6 mA typical

12 mA maximum

Service life with Alkaline IEC 6LR61 battery

Typical.............................. 55 hours 
Minimum........................... 40 hours

Battery indicator (ON)............... Green LED dims when battery 

voltage is <6.5 V

Safety

General..................................... IEC 61010-1: Pollution Degree 2
Measurement ........................... IEC 61010-2-032: CAT III 600 V

Electromagnetic Compatibility (EMC)

International.............................. IEC 61326-1: Portable 

Electromagnetic Environment
CISPR 11: Group 1, Class A 

Group 1: Equipment has intentionally generated and/or use 
conductively coupled radio-frequency energy which is necessary for 
the internal functioning of the equipment itself. 
Class A: Equipment is suitable for use in all establishments other 
than domestic and those directly connected to a low voltage power 
supply network which supplies buildings used for domestic purposes. 
There may be potential difficulties in ensuring electromagnetic 
compatibility in other environments, due to conducted and radiated 
disturbances. 
Caution: This equipment is not intended for use in residential 
environments and may not provide adequate protection to radio 
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur 
when the equipment is connected to a test object.

USA (FCC) ............................... 47 CFR 15 subpart B. This product is 

considered an exempt device per 
clause 15.103.

Electrical Specifications 

Valid temperature ............................. 23 °C ±3 °C (73 °F ±5 °F).
Current Ranges ................................ 0 A dc to 10 A dc or ac peak

0 A dc to 100 A dc or ac peak

Output Signals.................................. 10 A range: 100 mV/A

100 A range: 10 mV/A

Basic Accuracy (DC to 1kHz):

Input Current 

(DC or AC peak)

Error (after zero check)

Range 100 mV/A 

Range 10 mV/A

0 A to 10 A
0 A to 40 A
40 A to 80 A
80 A to 100 A

<3 % of r50 mA

-
-
-

-

<4 % of r50 mA
<12 % of r50 mA
<15 % of reading

80i-110s.book  Page 8  Monday, December 5, 2016  12:12 PM

Summary of Contents for 80i-110s

Page 1: ...m The 80i 110s is also compatible with any instrument capable of millivolt measurements The Probe provides these benefits Accurate AC DC and AC DC current measurements for Electrical Electronic and Au...

Page 2: ...read all instructions Do not use the Product if it is altered or damaged Do not alter the Product and use only as specified or the protection supplied by the Product can be compromised Comply with loc...

Page 3: ...ors is permitted Conforms to European Union directives Conforms to relevant Australian Safety and EMC standards Measurement Category III is applicable to test and measuring circuits connected to the d...

Page 4: ...he Probe is in the OFF position 3 Locate the battery cover on the handle and loosen the screw with a flat blade screwdriver 4 Slide the battery cover away from the Probe 5 Install the battery IEC 6LR6...

Page 5: ...NC to banana adapter PM9081 001 to connect the Probe to the input 2 On the Probe select the least sensitive range 10 mV A Make sure that the green ON indicator lights See Figure 4 for selector switch...

Page 6: ...rying conductors On the Probe the 100 mV A range offers the best accuracy Maintenance Before each use inspect the Probe Look for cracks or missing portions of the housing and output cable insulating c...

Page 7: ...p cloth and detergent Do not use abrasives or solvents Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth Do not allow rust or corrosion to form on the magnetic core ends If th...

Page 8: ...ic purposes There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Caution This equipment is not intended for us...

Page 9: ...perature from 0 C to 50 C 32 F to 132 F Performance Verification Verify Probe accuracy with the test setup shown in Figure 5 Required test equipment is defined in Table 1 Toroid coil construction is i...

Page 10: ...turns wound on cylindrical form using 10 gauge magnet wire Remove form and tape coil together making sure 10 wires are taped Tape D Fan out coil to approximately 270 with spacing 5 cm 2 in using tape...

Page 11: ...6 A 559 mV 641 mV Table 4 Performance Test Points Current Range 0 A to 100 A 10 mV A DC Measurement 5520A Settings DC Amps Measured Low Limit Output High Limit Output 1 A 10 A 95 5 mV 104 5 mV 3 A 30...

Page 12: ...uthorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE A...

Reviews: