Figure 5. Performance Test and Calibration Setup

Table 2. Toroid Coil Construction

Item

Description

10 turns wound on cylindrical form using 10-gauge magnet wire.

Remove form and tape coil together making sure 10 wires are 
taped.

 

Tape

D

Fan out coil to approximately 270° with spacing 5 cm (2 in) using 
tape to maintain coil shape.

5520A

DMM

10.5 cm (4 in)

D

80i-110s.book  Page 10  Monday, December 5, 2016  12:12 PM

Summary of Contents for 80i-110s

Page 1: ...m The 80i 110s is also compatible with any instrument capable of millivolt measurements The Probe provides these benefits Accurate AC DC and AC DC current measurements for Electrical Electronic and Au...

Page 2: ...read all instructions Do not use the Product if it is altered or damaged Do not alter the Product and use only as specified or the protection supplied by the Product can be compromised Comply with loc...

Page 3: ...ors is permitted Conforms to European Union directives Conforms to relevant Australian Safety and EMC standards Measurement Category III is applicable to test and measuring circuits connected to the d...

Page 4: ...he Probe is in the OFF position 3 Locate the battery cover on the handle and loosen the screw with a flat blade screwdriver 4 Slide the battery cover away from the Probe 5 Install the battery IEC 6LR6...

Page 5: ...NC to banana adapter PM9081 001 to connect the Probe to the input 2 On the Probe select the least sensitive range 10 mV A Make sure that the green ON indicator lights See Figure 4 for selector switch...

Page 6: ...rying conductors On the Probe the 100 mV A range offers the best accuracy Maintenance Before each use inspect the Probe Look for cracks or missing portions of the housing and output cable insulating c...

Page 7: ...p cloth and detergent Do not use abrasives or solvents Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth Do not allow rust or corrosion to form on the magnetic core ends If th...

Page 8: ...ic purposes There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Caution This equipment is not intended for us...

Page 9: ...perature from 0 C to 50 C 32 F to 132 F Performance Verification Verify Probe accuracy with the test setup shown in Figure 5 Required test equipment is defined in Table 1 Toroid coil construction is i...

Page 10: ...turns wound on cylindrical form using 10 gauge magnet wire Remove form and tape coil together making sure 10 wires are taped Tape D Fan out coil to approximately 270 with spacing 5 cm 2 in using tape...

Page 11: ...6 A 559 mV 641 mV Table 4 Performance Test Points Current Range 0 A to 100 A 10 mV A DC Measurement 5520A Settings DC Amps Measured Low Limit Output High Limit Output 1 A 10 A 95 5 mV 104 5 mV 3 A 30...

Page 12: ...uthorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE A...