7012 Calibration Bath
User’s Guide
54
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 9999.99 {C or F}
set: 105.00 C
Set current set-point to n
s[etpoint]=n
s=100
Instrument
Range
Read vernier
v[ernier]
v
v: 9.99999
v: 0.00000
Set vernier to n
v[ernier]=n
v=.00001
Depends on
Configuration
Read temperature
t[emperature]
t
t: 9999.99 {C or F}
t: 55.69 C
Read temperature units
u[nits]
u
u: x
u: c
Set temperature units:
u[nits]=c/f
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to
Fahrenheit
u[nits]=f
u=f
Secondary Menu
Read proportional band setting
pr[op-band]
pr
pr: 999.9
pr: 15.9
Set proportional band to
n
pr[op-band]=n
pr=8.83
Depends on
Configuration
Read cutout setting
c[utout]
c
c: 9999 {x},{xxx}
c: 620 C, in
Set cutout setting:
c[utout]=n/r[eset]
Set cutout to
n degrees
c[utout]=n
c=500
Temperature
Range
Reset cutout now
c[utout]=r[eset]
c=r
Read heater power
(duty cycle)
po[wer]
po
po: 9999
po: 1
±100
Configuration Menu
Probe Menu
Read D0 calibration parameter
*d0
*d0
d0: 999.9999
d0: -25.2290
Set D0 calibration parameter to
n *d0=n
*d0=-25.2290
-999.9999 to
999.9999
Read DG calibration parameter
*dg
*dg
dg: 999.9999
dg:186.9740
Set DG calibration parameter to
n *dg=n
*dg=186.9740
-999.9999 to
999.9999
Operating Parameters Menu
Read cutout mode
cm[ode]
cm
cm: {xxxx}
cm: AUTO
Set cutout mode:
cm[ode]=r[eset]/a[uto]
RESET or
AUTO
Set cutout to be reset manually-
cm[ode]=r[eset]
cm=r
Set cutout to be reset
automatically
cm[ode]=a[uto]
cm=a
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Table 4
Interface Command Summary.