4.3
Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability. Refer to Section 3.1, Specifications, for power require-
ment details. Refer to and read the WARNING at the front of this manual con-
cerning brownout and over voltage protection.
Set the “HEATER” switch on the front panel to position “LOW” and turn the
bath on using the front panel “POWER” switch. The bath will turn on and be-
gin to heat or cool to reach the previously programmed temperature set-point.
The front panel LED display will indicate the actual bath temperature.
4.4
Setting the Temperature
In the following discussion a solid box around the word SET, UP, EXIT or
DOWN indicates the panel button while the dotted box indicates the display
reading. Explanation of the button or display reading are to the right of each
button or display value.
To view or set the bath temperature set-point proceed as follows. The front
panel LED display normally shows the actual bath temperature.
24.68 C
Bath temperature display
When “SET” is pressed the display shows the set-point memory that is cur-
rently being used and its value. Eight set-point memories are available.
S
Access set-point selection
1. 25.0
Set-point 1, 25.0°C currently used
Press “SET” to select this memory and access the set-point value.
S
Access set-point value
C 25.00
Current value of set-point 1, 25.00°C
Press “UP” or “DOWN” to change the set-point value.
U
Increment display
C 30.00
New set-point value
Press “SET” to accept the new value and display the vernier value. The bath be-
gins heating or cooling to the new set-point.
7012 Calibration Bath
User’s Guide
14