background image

Table 3. Features and Components 

No. Item/Function 

 

Desiccant tube 

 

Pump shaft 

 

Inlet valve 

 

Vent valve 

 

Outer body end cap  

 

Handle connector 

 

Pump handle 

 

Variable volume 

 

Reference gauge connection port 

 

Calibration manifold 

 

Unit Under Test (UUT) connection port with 
replaceable filter 

 

Inlet port 

 

Pump bracket 

 

Pump connection port 

 

Pin 

 

O-Ring 

 

Ball bearings 

 

Large piston nut 

 

Shaft seal 

 

Pump shaft 

 

Ball/seal retainer nut black seal 

 

Piston seal (green) 

 

Ball/seal retainer nut 

Set Up 

700HPP (Pump only) 

1.  Take the pump out of the box.  
2.  Open the folding legs for stability. 
3.  Attach one end of the hose to the port on the pump 

base (hand tighten only). 

4.  Connect the other end of the hose to the Unit Under 

Test (UUT) with the standard fittings provided.  

5.  Before you generate pressure, make sure the vent 

valve located at the bottom of the pump next to the 

desiccant tube is completely closed.  

700HPPK (Pump kit) 

1.  Take the pump out of the box.  
2.  Open the folding legs for stability. 
3.  Attach the calibration manifold to the pump bracket as 

shown in Figure 1. The calibration manifold is 
detachable due to the quick detent pins.  

4.  Attach one end of the first hose to the port at the base 

of the pump (hand tighten only).  

5.  Wind the hose around the hooks of the bracket and 

attach the other end to the inlet port at the bottom of 
the calibration manifold (hand tighten only). 

6.  Attach the second hose to the UUT with the standard 

fittings provided and the other end to the UUT port on 
the calibration manifold (hand tighten only). 

7.  Install the reference gauge to the reference port of the 

calibration manifold. The reference gauge connection 

is a 1/4 NPT female connection that seals on an O-
ring. PTFE tape is not required. 

Operation 

The subsequent sections explain Product operation.  

700HPP Operation 

To operate the pump: 
1.  Attach the free end of the hose to the device to be 

pressurized with the fittings provided.  

2.  Make sure that the vent valve at the bottom of the 

Product is closed. 

3.  Pump the Product to generate pressure.  

For maximum efficiency, use full pump strokes. 
Generally, 21 MPa (3000 psi) can be generated within 
20 seconds when full strokes are used (for volumes of 
approximately 10 cm

3

). 

4.  To vent the pressure, use the vent valve on the base 

of the pump. 

5.  After use, wind the hose around the hooks for ease of 

use and portability. 

6.  Fold the pump base and place in the bag. 

700HPPK Operation 

To operate the pump and calibration manifold: 
1.  The 700HPPK can be operated with the calibration 

manifold attached as shown in Figure 1 or detached 
for easier operation. 

2.  Make sure the “inlet” valve is set to open on the 

calibration manifold and the two vent valves are firmly 
closed (at the base of the pump and on the calibration 
manifold marked “VENT”).  Attach the reference 
gauge to the port nearest to the top of the calibration 
manifold (reference gauge connection port).  

3.  Attach the UUT (external) line to the port towards the 

lower end of the calibration manifold.  

4.  For each pressure that you measure, make sure that 

the reference gauge shows zero pressure. If not, 
open the vent valve on the calibration manifold and 
zero the reference gauge manually.  

5.  Close the vent valve.  

6.  Pump the Product to generate pressure.  

For maximum efficiency, use full pump strokes. 
Generally, 21 MPa (3000 psi) can be generated within 
20 seconds when full strokes are used (for volumes of 
approximately 10 cm

3

). 

7.  Once the applied pressure comes near the targeted 

pressure, close the inlet valve on the calibration 
manifold. This will help to stabilize the pressure faster.  

8.  Use the black variable-volume knob on the calibration 

manifold to adjust the pressure and achieve exact 
target pressure (cardinal point) on the reference 
gauge.  

Note 

Do not use the gauge at the bottom of the pump 
for calibration. 

9.  Take readings from both reference and UUT pressure 

gauges for calibration.   

Summary of Contents for 700HPPK

Page 1: ...nder test Warning To prevent personal injury Read all safety information before you use the Product Use the Product only as specified or the protection supplied by the Product can be compromised Carefully read all instructions Do not use the Product if it operates incorrectly Do not use the Product if it is altered or damaged Disable the Product if it is damaged Do not attempt to operate the Produ...

Page 2: ...the Product after maintenance or repair 1 Wear eye protection and securely attach a pressure gauge to the hose or accessory fitting of the calibration manifold 700HPPK only Use any approved properly rated adapter as required 2 Close the upper vent valve 700HPPK and slowly pressurize the pump gauge system to 21 MPa 3000 psi 210 bar 3 Maintain pressure for 60 seconds 4 Use the upper vent valve on th...

Page 3: ...end of the hose to the device to be pressurized with the fittings provided 2 Make sure that the vent valve at the bottom of the Product is closed 3 Pump the Product to generate pressure For maximum efficiency use full pump strokes Generally 21 MPa 3000 psi can be generated within 20 seconds when full strokes are used for volumes of approximately 10 cm3 4 To vent the pressure use the vent valve on ...

Page 4: ...old slightly to drain all the liquid contaminants from inside that port 4 Clean the screen of any particulates or replace if necessary 5 Reinstall the screen and tighten the fitting Lubricate the Pump If generation of pressure becomes difficult due to friction in the pump then the pump may require lubrication Depending on usage lubrication of the pump may be required monthly Lack of lubrication ca...

Page 5: ...ere is a large black seal If it is worn replace it When you reassemble the pump add small amount of silicone grease to this seal 9 Remove the small diameter piston rod 10 On the bottom of the small diameter piston rod there is a small black seal If it is worn replace it The seal on top of the ball seal retainer nut should not need replacement 11 On the bottom of the piston there is a small green s...

Page 6: ...sable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a des...

Reviews: