Box Contents 

The Product ships with 

 

700HPP High Pressure Pneumatic Test Pump 

 

700HPM High Pressure Calibration Manifold with 

hose (700HPPK only) 

 

Canvas carrying bag 

 

High Pressure Pneumatic Quick Service Kit 

 

Dry pac refill 

 Silicone 

grease 

 

High Pressure hose  

 Instructions 

 

Fittings (as listed in Table 2) 

Table 2. Pump Fittings 

Configuration 

Ordered 

M16- ¼ 

NPT 

(Male) 

M16-

 

NPT 

(Male) 

M16-¼ 

BSP 

(Male) 

M16-

 

BSP 

(Male) 

M16-

M14 

(Male) 

¼ NPT 

(F)-

M20x 

1.5 

(Male) 

700HPP-
NPT  

1 1 - - - - 

700HPPK-
NPT 

700HPP-
BSP  

- - 1 1 - - 

700HPPK-
BSP 

700HPP-
MET  

1 - - - 1 1 

700HPPK-
MET 

Safe Operation 

To verify safe operation of the Product after maintenance 
or repair: 
1.  Wear eye protection and securely attach a pressure 

gauge to the hose or accessory fitting of the 

calibration manifold (700HPPK only). Use any 
approved (properly-rated) adapter as required.  

2.  Close the upper vent valve (700HPPK) and slowly 

pressurize the pump/gauge system to 21 MPa 
(3000 psi, 210 bar).  

3.  Maintain pressure for 60 seconds.  
4.  Use the upper vent valve on the calibration manifold 

(700HPPK only) to slowly release the pressure. 

Pressure can be vented using the vent valve at the 
base of the pump when using the 700HPP. 

Features 

The features and components of the Product are shown in 

Figure 1 and Table 3. 

A

B

3

4

5

6

7

8

9

10

11

12

13

A

B

14

15

1

2

16

23

21

22

17

18

19

20

 

hza01.eps 

Figure 1. Features and Components 

 

Summary of Contents for 700HPPK

Page 1: ...nder test Warning To prevent personal injury Read all safety information before you use the Product Use the Product only as specified or the protection supplied by the Product can be compromised Carefully read all instructions Do not use the Product if it operates incorrectly Do not use the Product if it is altered or damaged Disable the Product if it is damaged Do not attempt to operate the Produ...

Page 2: ...the Product after maintenance or repair 1 Wear eye protection and securely attach a pressure gauge to the hose or accessory fitting of the calibration manifold 700HPPK only Use any approved properly rated adapter as required 2 Close the upper vent valve 700HPPK and slowly pressurize the pump gauge system to 21 MPa 3000 psi 210 bar 3 Maintain pressure for 60 seconds 4 Use the upper vent valve on th...

Page 3: ...end of the hose to the device to be pressurized with the fittings provided 2 Make sure that the vent valve at the bottom of the Product is closed 3 Pump the Product to generate pressure For maximum efficiency use full pump strokes Generally 21 MPa 3000 psi can be generated within 20 seconds when full strokes are used for volumes of approximately 10 cm3 4 To vent the pressure use the vent valve on ...

Page 4: ...old slightly to drain all the liquid contaminants from inside that port 4 Clean the screen of any particulates or replace if necessary 5 Reinstall the screen and tighten the fitting Lubricate the Pump If generation of pressure becomes difficult due to friction in the pump then the pump may require lubrication Depending on usage lubrication of the pump may be required monthly Lack of lubrication ca...

Page 5: ...ere is a large black seal If it is worn replace it When you reassemble the pump add small amount of silicone grease to this seal 9 Remove the small diameter piston rod 10 On the bottom of the small diameter piston rod there is a small black seal If it is worn replace it The seal on top of the ball seal retainer nut should not need replacement 11 On the bottom of the piston there is a small green s...

Page 6: ...sable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a des...

Reviews: