Portable Calibration Baths
Commands
59
SOURce:SENSe:CALibration:PARameter<index> <numeric_value>
SOURce:SENSe:CALibration:PARameter<index>?
Set a calibration parameter for the control sensor.
<index> specifies the calibration parameter; 1: TEMP1; 2: TEMP2; 3: TEMP3.
<numeric_value> is the value to which the calibration parameter is set.
Calibration parameters should only be changed by a qualified technician as part
of a calibration procedure.
This command is password protected.
SOURce:SENSe:AVERage:[:AVERage]?
Return the moving average of the fluid temperature.
The response is a temperature value in
°
C or
°
F and depends on the temperature
unit setting.
SOURce:SENSe:AVERage:SDEViation?
Return the moving standard deviation of the fluid temperature.
The response is a temperature value in
°
C or
°
F and depends on the temperature
unit setting.
SOURce:SENSe:DATA?
Return the fluid temperature.
The response is a temperature value in
°
C or
°
F and depends on the temperature
unit setting.
SOURce:SENSe:ROUTe <boolean>|DEFault
SOURce:SENSe:ROUTe?
Select the control sensor source.
<boolean> is 0 (Internal sensor) or 1 (reference probe).
This command is conditionally password protected.