![Fluke 6109A Operator'S Manual Download Page 33](http://html1.mh-extra.com/html/fluke/6109a/6109a_operators-manual_2310363033.webp)
Portable Calibration Baths
Operation
29
The extension “.txt” is automatically appended to the filename when the file is
created. If the file already exists, new data is appended to the end of the file. The
file is placed in a folder determined by the serial number of the Product:
\Fluke\7109_6109\<product serial number>\Data\
Each line in the file contains the date and time, fluid temperature, its unit of
measurement, reference probe reading, its unit of measurement, DUT reading,
and its unit of measurement. Fields are separated by commas. If reference probe
or DUT readings are not available, the field is empty.
Run a Program
A program instructs the Product to step through a sequence of setpoints to
automatically test sensors at multiple temperatures.
To set up a program:
1. Push
.
2. Push
(
New
).
3. Push
to specify the program name. See
Keyboard Use
.
4. Push
(
Save
) to save the program name.
5. Push
(
Edit Setpoints
) to specify the setpoints.
6. Push
or
to choose an existing setpoint or
to add a new setpoint.
7. Push
to edit the setpoint temperature.
8. Use the number keys to type the setpoint temperature.
9. Push
to save the setpoint.
10. Push
after all setpoints have been defined.
11. Push
several times to move the highlight to Dwell Time.
12. Push
to edit the Dwell Time.
13. Use the number keys to type in a time in minutes.
14. Push
to save the value.
15. Push
(
Save
) to save the program.
If the Input Module (7109A-P or 6109A-P) is installed, the module can collect the
sensor readings and produce a test report. To turn on program report:
1. Push
.
2. Push
or
to choose a program.
3. Push
(
SELECT
).
4. Push
(
Options
).
5. Push
to move the highlight to Program Report.
6. Push
to change Program Report to ON.
7. Push
to move to Pass Tolerance.
8. Push
to change the number.
9. Use the number keys to type in the new number.
10. Push
to save the value.
www.
.com
1.800.561.8187