Fluke 55XXA Instructions Download Page 3

Portability Kit

Installation

3

Installation

When you install the Kit on the Product, do not remove too many screws from the Product at 

once. To do so can cause the Product covers to shift and make the Product come apart.
To install the Kit on the Product, See 

Figure 3

:

1. Turn off the Product and remove the mains power cord. 
2. Place the Product on a flat surface. 
3. Start with one side (left or right). Use a 3 mm hex wrench to remove the side front and rear 

handles (each handle has 5 screws). 

4. With the M4x10 screws, attach one of the handle assemblies to the Product. Use screw 

holes that held the standard handles. Use five screws for each handle. All 20 M4 handle 

screws require a Torx T20 drive. Tighten to a torque of 12.5 in-lbs. 

5. Attach the top-panel rear cross bar to the Product. See 

Figure 3

:

a. On the rear of the Product top-panel, use a Phillips screwdriver to remove the four 

screws. Do not remove the screws from the top-panel front side yet. 

b. With the slots of the cross bar facing out, slide one end tab of the cross bar into the 

hole on the corner handle. 

c. Use four M3X8 screws to attach the cross bar to the top rear of the Product. The four 

cross bar screws require a Phillips bit. Torque of 8 in-lbs. 

Note

Start each of these screws but wait to tighten until all are started.

Figure 3. Attach the Cross Bar to the Handle and Top of the Product

Cross Bar

Cross Bar

(4) M3X8 Screws