5320A-LOAD
High Voltage Load Adapter
4
Table 1. 5320-LOAD Resistance Verification Values
Voltage Method – Verifying the Load Resistances while under Power
Alternative DMM
Verification
Method
Nominal
Resistance
Value
Externally
Applied Test
Voltage at
mains
Frequency
Nominally
Required Test
Current
Measured
Current
Calculated
Resistance
RL=Vs/Im
(
±
10 % of nominal)
Measured
Resistance
(
±
10 % of nominal)
10 k
Ω
200 Vac
20 mA
35 k
Ω
315 Vac
9 mA
50 k
Ω
300 Vac
6 mA
100 k
Ω
600 Vac
6 mA
250 k
Ω
1000 Vac
4 mA
500 k
Ω
1000 Vac
2 mA
1 M
Ω
1000 Vac
1 mA
5 M
Ω
1000 Vac
0.2 mA
Replaceable Parts
Table 2 lists the replaceable parts for the Load. To contact Fluke, visit Fluke’s website at
www.fluke.com
or call one of the following numbers:
USA and Canada: 1-888-99-FLUKE (1-888-993-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
China: +86-10-6512-343
5-2
Singapore: +65 6799-5588
Anywhere in the world: +1-425-446-5500
Table 2. Replaceable Parts
Description Fluke
PN
POWER ADAPTER 100-240V AC, 12 V DC, with five plug adapters
3132484
GROUNDING CABLE
3132491
General Specifications
Power supply voltage
........................................... AC adapter 100-240 V, output voltage 12 V @ 0.4 amps min.
Warm-up time
........................................................ Not applicable
Specifications confidence level
.......................... 99 %
Temperature
Operating Temperature ...................................... 5
°
C to 40
°
C
Recommended Calibration Temperature (Tcal) . 23
°
C
Storage Temperature ......................................... -20
°
C to +70
°
C
Altitude, Maximum
Operating ............................................................ 3,050 m (10,000 ft)
Storage ............................................................... 12,200 m (40,000 ft)
Dimensions
........................................................... 430 mm X 462 mm X 95 mm (16.9 in X 18.2 in X 3.7 in)