High Voltage Load Adapter
Performance and Verification Test
3
Note
A Fluke 5520A or 5500A Calibrator is
recommended as a source for the voltage
method due to its output voltage/current
capabilities. A Fluke 8845A Digital
Multimeter (or equivalent) is
recommended for measurements with
either test method.
For the voltage test method, test each resistor as
follows:
1. Apply the recommended voltage at power
mains frequency (either 50 or 60 Hz)
between the terminal for the resistor being
measured and the COM terminal.
2. Measure the current flowing through the load.
3. Calculate the resistance by dividing the
sourced voltage by the measured current
(R
L
=V
s
/I
m
).
The calculated resistance should be within 10 %
of the nominal resistance value found in Table 1.
Repeat steps 1 through 3 for each Load resistor
terminal, adjusting the applied voltage per
Table 1.
For the alternative multimeter test method,
measure each resistor as follows:
1. Place the probes of a multimeter between the
selected resistor input terminal and the COM
terminal.
2. Read the measured resistor using the
resistance mode of the multimeter, and note
the value.
3. Insure the measured value is within 10 % of
the nominal resistance.
Repeat steps 1 through 3 for each Load resistor
terminal per Table 1.
POWER
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
EXGRD
SCOPE
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
÷
MODES
MORE
STBY
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
5520A
CALIBRATOR
SCOPE
OUT
V, , ,RTD
A, -SENSE, AUX V
20V PK MAX
20V PK MAX
CALIBRATION
ENABLE
SERIAL 1
FROM HOST
IEEE-488
IN OUT
/
63 Hz
47 Hz
600VA MAX
FLUKE CORPORATION
MADE IN USA
PATENTS PENDING
NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
NORMAL
WARNING:
TO AVOID ELECTRIC SHOCK GROUNDING
CONNECTOR IN POWER CORD MUST BE CONNECTED
FUSE
CAUTION
FOR FIRE PROTECTION REPLACE ONLY
WITH A 250V FUSE OF INDICATED RATING
MAINS SUPPLY
100V
/
120V
220V
/
240V
T5.0A 250V (SB)
T2.5A 250V (SB)
CHASSIS
GROUND
SERIAL 2
TO UUT
10 MHz
MAX
5V P - P
WARNING:
TO AVOID PHYSICAL INJURY, INSURE THAT THE FILTER
IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRUMENT
TO CLEAN THE FILTER:
-UNPLUG INSTRUMENT
-REMOVE FILTER
-FLUSH WITH SOAPY WATER
-DRY BEFORE REINSTALLATION
INSTALLED
OPTIONS
- SC300
- SC600
5320A-LOAD
1200 V MAX
140 W MAX
2000 V MAX
110 W MAX
2000 V MAX
80 W MAX
5500 V MAX
300 W MAX
5500 V MAX
120 W MAX
5500 V MAX
60 W MAX
5500 V MAX
30 W MAX
5500 V MAX
6 W MAX
COM
10 k
35 k
500 k
1M
50 k
100 k
250 k
5 M
100 mA
40 mA
10 mA
4 mA
1mA
POWER
I
O
20 V MAX
PROTECTIVE
EARTH (GROUND)
POWER INPUT
12 V
10 VA MAX.
NO USER SERVICEABLE PARTS CONTAINED WITHIN.
HAZARDOUS VOLTAGES PRESENT. DO NOT REMOVE COVERS.
SAFETY WARNING
DISCONNECT POWER AND SIGNAL LEADS
BEFORE SERVICING.
READ MANUAL BEFORE USE.
REFER SERVICE TO QUALIFIED PERSONNEL.
FLUKE CORPORATION
www.fluke.com
DURING USE,
TERMINAL MUST BE
CONNECTED TO THE
PROTECTIVE EARTH
OF THE 5320A
8846A
6-1/2 DIGIT PRECISION MULTIMETER
0.20100
mA~
DC V
ACV
DC I
AC I
TRIG
FREQ
PERIOD
TEMP
ZERO
INSTR
SETUP
MEAS
SETUP
RANGE
F1
F2
F3
F4
F5
BACK
MEMORY
ANALYZE
INPUT
SENSE
HI
LO
FRONT
REAR
10A
1V
400
mA
4W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
Front
Rear
ehq104.eps
Figure 4. Test Equipment Connections for Alternate Test Method