Remote Operations
Remote Commands
3
3-11
6105A SCPI Command Details
This section contains the syntax rules for 6105A SCPI commands that control the Product
when it is controlled by the 6105A.
[:SOURce]:PHASe<x>:CURRent:EAMP:RANGe(?) <nrf>, <nrf>
Description
Sets or gets the current range of the Product assigned to phase identified by
x.
Parameters
<nrf>,<nrf> = 0,2
Sets output current range to 2 A.
= 0,20
Sets output current range to 20 A.
= 0,120
Sets output current range to 120 A.
Example
PHAS1:CURR:EAMP:RANG 0,20
Sets output current range to 20 A.
PHAS1:CURR:EAMP:RANG?
Gets
back
0,2
,
0,20
, or
0,120
.
[:SOURce] PHASe<x>::CURRent:EAMP:LCOMp(?) <bool>
Description
Sets or retrieves the state of the LCOMP (inductive-load compensation)
function.
Parameters
<bool> = ON or 1
Turn on LCOMP.
OFF or 0
Turns off LCOMP.
Example
PHAS1:CURR:EAMP:LCOM ON
Turns on LCOMP on the Product set to
phase 1.
PHAS:CURR:EAMP:LCOM?
Gets
back
1
or
0
.
[:SOURce] PHASe<x>::CURRent:EAMP:TERMinal:MODE(?) <cpd>
Description
Sets or gets the state of which output terminals are used for different
ranges.
Parameters
<cpd> = HIGH
Output all ranges on the high terminals.
AUTO
Output the 2 A and 20 A range on the low
terminals and output the 120 A range on the
high
terminals.
Example
PHAS1:CURR:EAMP:TERM:MODE HIGH
Output all ranges on the high
terminals.
[:SOURce] PHASe<x>:CURRent:EAMP:TERMinal:ROUTe?
Description
A query only command that gets back which front-panel output terminal is
in use.
Example
PHAS1:CURR:EAMP:TERM:ROUT?
Gets
back
HIGH
or
LOW
.
[SOURce]:PHASe<x>:CURRent:EAMP:FITTed?
Description
Gets back the number of units in the chain for the phase identified by x.
Example
PHAS1:PHAS:CURR:EAMP:FITT?
Gets back a number from 1 to n.
[SOURce]:PHASe<x>:CURRent:EAMP:IDN? <nrf>
Desription
Gets the identification string of an instrument in the chain of the phase
identified by x.
Parameter
<nrf>
= 1 to 10
Index of the instrument in the chain.
Example
PHAS1:CURR:EAMP:IDN? 2
Gets back the identification string of the
second instrument in the chain in the
phase identified by x.
Returns
<manufacturer>, <model>, <serial number>, <firmware>
Summary of Contents for 52120A
Page 2: ......
Page 6: ...52120A Users Manual iv...
Page 8: ...52120A Users Manual vi...
Page 10: ...52120A Users Manual viii...
Page 12: ...52120A Users Manual 1 2...
Page 32: ...52120A Users Manual 1 22...
Page 34: ...52120A Operators Manual 2 2...
Page 52: ...52120A Operators Manual 2 20...
Page 54: ...52120A Users Manual 3 2...
Page 66: ...52120A Users Manual 3 14...
Page 68: ...52120A Users Manual 4 2...
Page 72: ...52120A Users Manual 5 2...