52120A
Users Manual
1-8
To increase the life of the Product:
•
Keep the area around the air filter a minimum of 4 inches from nearby walls or
rack enclosures.
•
Keep the inlet and exhaust perforations on the sides of the Product clear of
blockages.
•
Keep the air that goes in to the Product between 5
°
C and 35
°
C.
•
Make sure exhaust from a different instrument is not pointed into the fan inlet.
•
Clean the air filter at a maximum of 30 day intervals. More frequently if the
Product is operated in a dusty environment. See the Maintenance section of this
manual for instructions on how to clean the air filter.
How to Connect the Product to Mains Power
XW
Warning
To prevent possible electrical shock, fire, or personal injury:
•
Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
•
Replace the mains power cord if the insulation is damaged
or if the insulation shows signs of wear.
•
Make sure the ground conductor in the mains power cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.
•
Do not disconnect or open the protective ground conductor
inside or outside the Product. An open ground conductor
can make the Product dangerous.
W
Caution
When exposed to low temperatures for an extended time, such
as air travel or storage, condensation may form inside the
Product. To prevent damage to the Product, let it acclimate to
its environment out of its shipping container a minimum of one
hour before you connect it to mains power.
The Product automatically senses mains voltage between 100 and 240 volts. No line
voltage or fuse selection is necessary. See the Maintenance section to learn how to
replace the mains fuse.
Because the Product can pull more current than a standard 10 A IEC connector, the
Product has a 16 A power connector on the rear panel. A mains power cord with 16 A
capacity is also supplied with the Product. Table 1-3 is a list of power cord types
available from Fluke.
Summary of Contents for 52120A
Page 2: ......
Page 6: ...52120A Users Manual iv...
Page 8: ...52120A Users Manual vi...
Page 10: ...52120A Users Manual viii...
Page 12: ...52120A Users Manual 1 2...
Page 32: ...52120A Users Manual 1 22...
Page 34: ...52120A Operators Manual 2 2...
Page 52: ...52120A Operators Manual 2 20...
Page 54: ...52120A Users Manual 3 2...
Page 66: ...52120A Users Manual 3 14...
Page 68: ...52120A Users Manual 4 2...
Page 72: ...52120A Users Manual 5 2...