51 & 52 Series II
Users Manual
10
Changing a Setup Option
1. Press
K
or
N
to scroll to the setup option you
want to change.
2. Press
E
to indicate that you want to change this
setting.
3. Press
K
or
N
until the setting you want to use
appears on the display.
4. Press
E
to store the new setting in memory.
Notes
Setup is disabled in MIN MAX mode.
Offset:
The primary display shows the temperature plus
the offset and the secondary display shows the
offset. Remember to reset the offset to 0.0 when
it is no longer needed. The offset automatically
resets to 0.0 when you change the thermocouple
type. Model 52: You can store individual offsets
for T1 and T2.
Sleep mode:
The thermometer enters sleep mode if no button
press occurs for 20 minutes. Pressing any button
wakes the thermometer and returns it to its
previous state. Sleep mode becomes enabled
each time you turn on the thermometer and is
automatically disabled in MIN MAX mode.
Line frequency:
For optimum rejection of line noise, the
thermometer must be set for the local line
frequency.