51 & 52 Series II
Replacement Parts and Accessories
15
Electrical (cont.)
Measurement
Accuracy, T1,
T2, or T1-T2
(Model 52)
J-, K-, T-, and E-type:
±
[0.05
%
of
reading
+
0.3
o
C (0.5
o
F)]
[below
−
100
o
C (
−
148
o
F): add 0.15 %
of reading for J-, K-, E-, and N-type;
and 0.45 % of reading for T-type]
Temperature
Coefficient
0.01
%
of reading
+
0.03
o
C per
o
C
(0.05
o
F per
o
F) outside the
specified
+
18
o
C to 28
o
C (
+
64
o
F
to
+
82
o
F) range
[below
−
100
o
C (
−
148
o
F): add
0.04 % of reading for J-, K-, E-, and
N-type; and 0.08 % of reading for
T-type]
Electromagnetic
Compatibility
Susceptibility:
±
2
o
C (
±
3.6
o
F) for 80
MHz to 200 MHz in 1.5 V/m field, for
200 MHz to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1
Maximum
Differential
Common Mode
Voltage
1 V (Maximum voltage difference
between T1 and T2)
Temperature
Scale
ITS-90
Applicable
Standards
NIST-175
Accuracy is specified for ambient temperatures
between 18
o
C (64
o
F) and 28
o
C (82
o
F) for a period of
1 year. The above specifications do not include
thermocouple error.
Replacement Parts and Accessories
Accessory
Part Number
Holster and Flex Stand
Assembly
1272438
AA NEDA 15A IEC LR6 batteries
376756
80PK-1 K-Type Bead Thermocouple
773135
CD-ROM
1276106
Service Manual
1276123