24
Measuring the Load on a Transformer
Measure the total kVA on all three phases to check the load on a transformer.
1
Open the main menu.
2
3
Make connections as shown below:
Look at the kVA reading. It shows the apparent power on phase 1. Write
down the value (kVA
1
).
The symbol of a Capacitor or an Inductor is shown to indicate capacitive or
inductive loads.
4
Repeat the measurement on phase 2 and phase 3 (keep the black test
lead connected to the neutral). Write down the values for kVA
2
and
kVA
3
and calculate kVA
TOTAL
_____ kVA
1
+ _____ kVA
2
+ _____ kVA
3
= _____ kVA
TOTAL
Compare this result with the transformer kVA rating. If the result is close to, or
over the nameplate reading of the transformer, reduce the load on the